NVT-E
Operating/Assembly instructions
KTR-N
Sheet:
Edition:
42611 EN
22 of 27
4
Please observe protection
note ISO 16016.
Drawn:
2019-12-06 Pz/Bet
Replacing:
KTR-N dated 2019-02-14
Verified:
2019-12-06 Pz
Replaced by:
The analogue output can be tested, too. The highest, the average and the smallest analogue value can be
displayed one after another. The setting is made in the following menu:
Test for Analog X
Check AnX
C)AN_
n)oP - FULL
Parameter
____
Settings Analog X
Analog OutX
An_
►
►
Main menu meas. value
Meas. value
____
►
The following menu items are available:
Normal operation
Output of highest analogue value
Output of average analogue value
Output of lowest analogue value
On completion of the tests make absolutely sure to set the function to normal operation
.
The device is in a position to record the details for a switching output. A result is defined as an illumination of the
LED. Accordingly, recording of the switching processes depends on the setting of the switching function of LEDs.
The settings and the evaluation can be performed as follows:
Measured values
Display
____
Menu diagnostics
Diagnostic
di A
►
It is only possible to record one switching output. The switching output to be recorded is
set in the menu item switching output Log Alarm
.
Press the button
to open the main menu.
Select the menu item
via the buttons
and
.
Please note:
This menu provides various diagnosis values and recordings for monitoring of filling
level and temperature.
•
Open the menu via the button
.
Now you are in a position to amend or retrieve the settings for diagnosis.
4
Operation
4.8 Analogue outputs
4.8.4 Testing the analogue output
4.9 Diagnosis