Error state cannot
be released.
Load had been
applied when the
power was turned on.
Temperature
changed suddenly
when the power was
turned on.
Measurement was
taken while a point
other than the grip
was being held.
No measurement
can be made.
The handle is in the
Peak Hold mode.
The handle is in the
setting state of Preset
mode.
Cancel the measured value by
pressing the +/C button.
Turn the power off and on again in
a place where the temperature is
close to the ambient temperature of
the operating environment.
Reading is
unreliable and
has significant
error.
Excessive force was
applied on the grip
during measurement.
All values other than N-m
are reference values.
Registered value and set value
are contradicting each other.
Battery has run out.
Indicator flashes to denote a
reference value.
Unit
M
Display is
flashing.
Cause
Trouble
• Selector lever : Make sure that the selector lever is operated properly and not
positioned halfway. Otherwise, the gear inside the mechanism will be
damaged, making the union mechanism ineffective. If the union
mechanism is made defective from any damage on the gear or
another inner part, replace the damaged part.
No. BR3E-K: Inner parts for No. BR3E (9.5sq ratchet handle)
PUSH
The ball is pressed against the
wall by the bush pin to hold the
socket wrench.
Selector lever
Release button
Match the registered and set
values (see -(4)).
9.
Union Mechanism
•
Union mechanism : The union mechanism ensures that the socket wrench does not
come off unless the release button is pressed. (Press the release
button also when inserting a socket wrench.)
Push pin
Ball
The ball is fit into the recess of the
push pin to release the socket
wrench.
10. Troubleshooting
Remedy
Instrument
cannot be
powered.
Insulating sheet has
not been removed.
Battery has run out.
Pull out the insulating sheet
(see "Preparation").
Whole
area
Replace the battery (see ).
7
3
Replace the battery (see ).
7
Apply a force on the grip slowly
when taking measurements
(see -(2)).
1
Do not apply any force on the grip
when turning the power on
(see -(1)).
1
Apply a force to the input point of
the grip (see -(2)).
1
Wait for 3 seconds for the handle to
enter the measurement state
(see 3 -(3)).
3
Refer to the error list (see ).
8