Maintenance
page 95
5. The next step is to enter the KRÜSS Tensiometer (K11 or K100) which is
connected:
Enter the instrument code <1=K11; 2=K100>
. Enter the number 1
for the K11 and press the ENTER key of the computer keyboard.
6. You are now asked to restart the K11 by pressing the Reset button on the rear
panel of the instrument:
Please press the reset button at the measuring
device <Ctrl-Break to abort>.
7. Carry out this procedure. The instrument now starts to initialise itself. The firmware
is then transferred from the PC to the K11.
*
If you wish to abort the update process at this point you should press the key
combination C
TRL
+B
REAK
.
In this case the window closes and the firmware on
your K11 remains unchanged.
8. When software transfer has finished the following window appears: Click
OK
. The
K11 is restarted and now contains the latest firmware version.