6.3 Wiring
The D3S Static Node contains HAZARDOUS LIVE parts.
The device must be installed with an external breaker so that
power can be disconnected as required. It is recommended
that the circuit breaker be installed near to the equipment.
The equipment should be positioned such that there is no
difficulty in disconnecting the D3S Static Node
The D3S Static Node contains HAZARDOUS LIVE parts.
External power must be disconnected before opening the lid
as this could expose HAZARDOUS LIVE parts.
6.3.1
Recommended Wiring
The wiring specification below is designed for standard single phase 16A Mains
supply. If the device is to be installed in a situation that does not have 16A current
limitation, then a current limiting device must be installed rated to 16A.
Number of Cores
3
Voltage Rating
450V
Current Rating
20A
Cross Sectional Area
1.5mm
2
Outer Diameter
9.2mm
American Wire Gauge
15
Example Supply
RS Components 744-0981
The wiring specification above is designed for standard single
phase 16A Mains supply. If the device is to be installed in a
situation that does not have 16A current limitation, then a
current limiting device must also be installed rated to 16A.
6.3.2
Anchor Points
The D3S Static Node contains HAZARDOUS LIVE parts.
Follow all instructions outlined in the Installation section of this
manual. Special care must be used to use the anchor points
for the mains cable to prevent risk of electrical shock.