Portable spectrum analyzer Arinst SSA-TG R2 6 April 2019
www.arinst.net | www.arinst.ru | www.kroks.ru
46
9.
EXAMPLES OF USE OF THE DEVICE
9.1.
Setting the internal oscillator output frequency and power
9.1.1. From the main menu of the control program, enter the
Generator
menu, as shown in figure 67.
igure 67 The setup menu parameters of the internal signal generator
9.1.2. Connect the generator output (1)
TG OUT
of the instrument and the antenna input (2)
RF IN
.
Depending on the characteristics of the device under test, set the oscillator frequency in the range from
35 to 6200 MHz, for example 1000 MHz, and the
Generator power
value in the range from -15 to -25
dBm.
In our example, the generator power value is -22 dBm.
If necessary, in the
Amplitude
settings menu, set the maximum value of the vertical scale
Ref level
in
the range from 10 to -20 dBm.
9.1.3. In the setup menu frequency characteristics
Frequency
select the frequency range of the over-
view
Span
and center frequency
Center
.
9.1.4. On the screen of your tablet/smartphone, will display a graph of the spectrum set the output of
the internal oscillator of the device, figure 68.