Data transfer
For the transfer of information from the flowmeter to the PC
proceed as follows:
-
Connect the RS 232 cable to the flowmeter and the serial
in/output port of the PC (com1 or com2).
-
Enter the programming mode of the flowmeter and
activate the function 3.6.1 or 3.6.2 for either the output of
the parameter settings or the measuring data; see that
the selection is on YES
-
Run program GETFLOW on the PC
For example if the Com1 port is connected and the data
has to be stored in file FLOW.DAT then type:
GETFLOW 1 FLOW.DAT
The message WAITING FOR DATA will appear.
-
Now press on the flowmeter within 3 seconds.
-
The data is transmitted to the PC and stored in the file
indicated using the ASCII format.
Data conversion
After all flow information is stored in an ASCII file use pro-
gram FLOW2CEL to convert the data to a format that can be
adapted for presentation purposes.
Run program FLOW2CEL on the PC and follow the instruc-
tions displayed.
See README _ E.DOC for any additional information.
C.4.3
Terminal programs
Standard terminal programs (e.g. MS Windows® terminal)
can be used to receive data from the UFC 600 T signal
converter by using a straight RS 232 cable with 25 pins
female D-connector.
Settings must be:
BAUDRATE
2400
DATA BITS
7
STOP BITS
1
PARITY
NONE