2.5 Protection type for gas hazardous areas
In type of protection "Intrinsic Safety" the sensor meets the requirements of EN 60079-11. The
explosion protection is ensured by limitation of the current and voltage so that no ignitable
energy can occur.
The marking of the sensor is:
II 1G Ex ia IIC T6...T4 Ga
2.6 Temperature classes
The permissible temperature range depending on the temperature class of the sensor is:
The maximum permissible temperatures are valid under the following conditions:
•
The sensor is operated in its intended mounting position.
•
The sensor is not exposed to heat radiation (e.g. direct sunlight, adjoining hot parts).
•
Insulation does not obstruct the ventilation of the sensor.
Components of the Ex marking and their definition
II
Group II explosion protection
1
Equipment category 1
G
Gas explosion protection
Ex ia
Intrinsically safe equipment, protection level ia
IIC
Gas groups IIA, IIB, IIC approved
T6...T4
Temperature class range T6...T4
Ga
EPL Ga approved, suitable in zone 0, zone 1, zone 2
Temperature
class
Permissible ambient temperature in
[°C]
Permissible product temperature in
[°C]
T4
-10...+80°C
0...+130°C
T6
-10...+40°C
0...+85°C
Temperature
class
Permissible ambient temperature in
[°F]
Permissible product temperature in
[°F]
T4
14...+176°F
32...+266°F
T6
14...+104°F
32...+185°F
INFORMATION!
The sensor has several temperature class information due to different ambient temperature
ranges. In principle, operation is possible in all ranges of the temperature classes T1...T6.