DESCRIPTION OF HART INTERFACE
42
SMARTPAT COND 1200
www.krohne.com
09/2015 - 4003041501 - MA SMARTPAT COND1200 R01 en
4 Device
1 Information
(read only)
1 Sensor information
1 Order code
2 Device name
3 Serial number
4 HART
®
ID
5 Polling address
6 Manufacturer ID
7 Date of manufacturing
8 SW version
9 HW version
2 Calibration
1 Measurand
2 Cell constant
3 Calibration solution
4 No. of calibrations
5 SIP counter
6 CIP counter
3 Operating parameters (OP) 1 Commissioning date
2 Operating hours
3 OP time > max OP t.
4 Max. OP temperature
5 Max. temperature
1
6 Temp. compensation
7 Temp. coefficient
2
8 Reference temp
2
2 Commissioning
Set date of commissioning
1
- Maximum reached temperature.
2
- Displayed only if linear temperature compensation is selected.
5 Service
1 Service calibration
1 Trimming at 4 mA (+/-)
2 Trimming at 20 mA (+/-)
2 Service parameter
1 Reset (Reboot sensor)
1 No
2 Yes
2 Load factory settings
(If you select Yes
Yes
Yes
Yes, the following settings will be reset to
default: temperature unit, measurand (PV), I/O, calibration
interval, 4...20 mA trimming, TAG and Long TAG. Polling
address will be set to 0 and loop current will be enabled.
1 No
2 Yes
3 Sensor lock
1 Lock sensor
If you select yes
yes
yes
yes the setting is not reversible anymore and
the sensor is invalid for use. The HART
®
communication is
not possible anymore.
1 No
2 Yes