5.4 Status messages and diagnostic information
Measurements out of specification
Maintenance
Message
Message
Message
Message
Description DD
Description DD
Description DD
Description DD
Description DTM
Description DTM
Description DTM
Description DTM
Action
Action
Action
Action
S01
Cond. > Max. cond.
Conductivity > Max. conductivity
Consider the measuring range
limits, otherwise select a suitable
sensor for the process conditions of
the application.
Conductive value out of range
S02
Cond. < Min. cond.
Conductivity < Min. conductivity
Conductive value out of range
S03
Res. > Max. res.
Resistivity > Max. resistivity
Resistivity value out of range
S04
Res. < Min. res.
Resistivity < Min. resistivity
Resistivity value out of range
S05
T > Temp. Max.
Temperature > Max. temperature
Temperature value out of range
S06
T < Temp. Min.
Temperature < Min. temperature
Temperature value out of range
Message
Message
Message
Message
Description DD
Description DD
Description DD
Description DD
Description DTM
Description DTM
Description DTM
Description DTM
Description DTM
Description DTM
Description DTM
Description DTM
M01
Data out of range
Calibration data out of range
Recalibrate the sensor. For more
information refer to
Calibration with
HART
®
Handheld 475 FIELD
COMMUNICATOR on page 25 or
refer to
Calibration with
PACTwareTM
on page 23.
Failure in calibration; cell constant deviation
M02
Maint. interval
Maintenance interval expired
Maintenance interval expired