5.1 Overview verification process
Step
(navigation tab)
Description
Tablet/laptop
start up OPTICHECK software by double clicking the desktop symbol
Start
converter is powered on
selected at startup of the software
; OPTICHECK shows a picture how to connect
connect tablet/laptop to the verification box and press "Connect"
select a flowmeter from the pictures; OPTICHECK shows picture
how/where to open housing and connect GDC
open housing and connect the verification box to the communication
interface (GDC) of the converter
press "Connect" ; the flow meter is identified
OPTICHECK takes the initial installation condition, which is used to
verify the reconnection at the end
OPTICHECK reads the identification data of the flowmeter
Configuration
select converter properties which cannot be automatically detected
select the verification level
select process conditions
(manually selection for some converters types)
all these data have influence on the verification
all these data have influence on the verification
all these data have influence on the verification
all these data have influence on the verification
OPTICHECK shows the instructions to connect the cables between the
verification box and the flow meter
converter power has to be switched off (to connect all cables) in most
cases
connect all cables
power on converter
Application
add information about the verification
Inspector name and Device location are mandatory fields
the information is stored with the verification data and printed on the
certificate
all these data have no influence on the verification
all these data have no influence on the verification
all these data have no influence on the verification
all these data have no influence on the verification
press "Apply"