Interpretation of verification results  

Check in the trend analysis which parameters are exceeding their limits. Most common causes are 
outlined below. 

 

5.1 

Accuracy data ADC, field current, mA or frequency output 

 
ADC is the primary signal processing. It includes input instrumentation amplifier and analogue to 
digital conversion of the flow signal voltage. Accuracy of ADC and field current directly influence 
accuracy of any output of the electromagnetic flowmeter. If accuracy of ADC, mA or frequency 
output, or field current is wrong: exchange converter electronics. 
 

5.2 Coil 

resistance 

 
Coil resistance value is dependent of coil temperature during verification. Coil temperature again is 
dependent on ambient temperature and on medium temperature in the flow tube of the magnetic 
flowmeter. Temperature coefficient is 0.4% per °C (0.2% per °F). Two failure modes can be 
detected: 

  Resistance is zero or far below plausible value: Check cabling, connectors, terminals to field 

coils on short circuits.  

  Resistance is high: Check cabling, connectors, terminals to field coils on interruptions or poor 

contact. 

Check connectors / connections below the terminal board. If these are ok and coil resistance 
remains infinite or zero, interruption or short-circuit of an internal connection is possible (but 
extremely rare).  In that case, the flowhead must be exchanged. 
 

5.3 

Electrode resistance with full pipe 

 
In general: Electrode resistance with full pipe depends on conductivity (

σ

) of the liquid. Conductivity 

depends on liquid type, concentration, temperature. Change of electrode resistance between two 
verifications only is significant as long as precisely the same liquid with same concentration and 
temperature will flow through the flowmeter without creating any deposits. Electrode resistance  
values may be estimated from electrode diameter “d”  and conductivity (

σ

) according to following 

formula: 

[ ]

[ ] [

]

100%

  / 

50%

:

Tol.

/

1000

+

×

=

cm

S

cm

d

k

R

Electrode

µ

σ

 

 
This formula may be used for approximate plausibility checks if the results of electrode resistance 
measurements are outside of the measuring range of MagCheck. The results of this formula may 
show a deviation to true value of electrode resistance as indicated by the tolerance data. 
 
Electrode resistance with full pipe very high: Check cabling to electrodes (signal cable, connector in 
remote flowhead connection box) on interruption. Remove meter from line, clean electrodes and 
grounding rings. 
 
Asymmetric electrode resistance: Differences in electrode 1- and 2 resistance values by more than 

±

50% may be caused by:  

  Heavy coating (contamination) on one of the electrodes (Countermeasure: Clean electrodes) 

  Beginning short circuit of one electrode, e.g. by contamination with highly conductive matter 

(Action: Clean electrodes and liner carefully) or by electrode leakage.  

 
Electrode leakage and conductive deposits can be confirmed by verification with empty pipe. 
Select “Pipe Empty” in Menu 1.1. 

 
 

Содержание MagCheck

Страница 1: ......

Страница 2: ...lead to loss of warranty In addition the general conditions of sale are applicable and are the basis for the purchase contract If the instrument needs to be returned to KROHNE please note the informat...

Страница 3: ...ual flow signal calibrator 14 3 4 MagCheck Menu functions for manual calibration test 17 4 Using MagCheck with a PC 21 4 1 System specifications 21 4 2 Installation PC software 21 4 3 Settings of MagC...

Страница 4: ...ersions Verification of the conventional electromagnetic flowmeter part of TIDALFLUX IFM 4110 PF is possible after disabling the level input to IFC 110 PF This will allow for a statement of flow veloc...

Страница 5: ...librator mode no data will be stored in MagCheck The measured indicated data must be noted and evaluated manually see Appendix 1 With non IMoCom converters In the flow signal calibrator mode with thes...

Страница 6: ...ow during the verification the meter Before changing any connections act according all process related safety and information rules existing in your organisation Disable all alarms and take all contro...

Страница 7: ...plied with MagCheck is only needed during download of verification data into a PC 2 3 Connection diagrams For use of MagCheck as precision flow signal calibrator for detailed manual verification of co...

Страница 8: ...8 MagCheck Connection MagCheck IFC 090 Connection MagCheck IFC 110 F...

Страница 9: ...EEPROM data for their validity After it will signal OK in case of an error it will indicate ERROR for approximately 5 seconds 2 5 1 With IMoCom converters IMoCom interface connected Next step is autom...

Страница 10: ...inus Correct field current is switching from positive to negative current and vice versa Plus Positive DC or pulsating DC defective converter negative half wave missing Minus Negative DC or pulsating...

Страница 11: ...hanging to submenus 3 1 l a n g u a g e To submenu l a n g u a g e E n g l i s h G B I M O C o m B u s F r e n c h F f i e l d c u r r e n t Back to main menu Ge r m a n D 2 6 3 Initial settings of Ma...

Страница 12: ...ro setting 1 4 Simulation Generation of flow signals adjustable in 0 1 steps of full scale value 1 5 Curr output Display of true value IMoCom converters only and actual value 1 6 Pulse output Display...

Страница 13: ...r at download into PC You may use up to 16 characters Next screen measurement Test conditions must be defined pipe full filled with liquid Electrode resistance will be tested via liquid With empty pip...

Страница 14: ...the GK will be reprogrammed automatically by MagCheck to original data 2 mA output tests verify compliance of mA output with factory calibration settings Time constant T settings via IMoCom is reduced...

Страница 15: ...e act value mA 1 6 Pulse output 1 6 Pulse output Shows pulse output at flow as entered under 1 4 True value Hz Actual value Hz Note true value Hz Note act value Hz 1 7 Field current 1 7 Field current...

Страница 16: ...l T900 000 1 3 Zero set 1 3 Zero set Allows MagCheck zero adjustment Physical unit as set under 1 2 Default 000 000 For standard tests use default For precision tests follow 3 4 2 1 4 Simulation 1 4 S...

Страница 17: ...ximum and minimum full scale values are subject to nominal diameter DN and the flow velocity v These limits are given in KROHNE documentation e g Installation and Operating Instructions of related con...

Страница 18: ...en 999 9 and 999 9 of the full scale as set or read under Function 1 2 Full Scale Value For normal tests setting shall be within 100 0 If the maximum signal level of MagCheck is exceeded the warning e...

Страница 19: ...hed Hardware Information and Status Information acc to menu 2 2 of IFC 010 020 090 110 are shown Display of Software I D no and status message of the connected module In this menu an 8 position softwa...

Страница 20: ...e supplied with MagCheck must be connected to the MagCheck connector the 9 pole connector to RS232 COM port of the PC COM port On the PC the MagCheck PC program must be active During transfer of the d...

Страница 21: ...for a password To work without password leave the field blank and click ok There is no way to recover a lost password After installation and clicking on the MagCheck icon on the desktop the main scre...

Страница 22: ...indicate Ready for data transfer Click on MagCheck PC main menu button Read flowmeter verification files from MagCheck MagCheck PC program will ask Number of new flowmeter verification files X e g 1...

Страница 23: ...ox allows for entering an appropriate file name e g MagCh_Date_All_UserComp MCK The list below shows all verification data files stored in file MagCheck MCK in the MagCheck data base Deselect files wh...

Страница 24: ...nd all settings as read by MagCheck Pull down lists will allow selection of a sequence of meter verifications in the meter history As an example the following view shows meter FT 1 parameter ADC 100 f...

Страница 25: ...selection menu This menu allows for selection of the device the parameter sequence to be printed print preview on the monitor printing as Adobe Acrobat Reader PDF file which may be stored or sent by...

Страница 26: ...ged 5 3 Electrode resistance with full pipe In general Electrode resistance with full pipe depends on conductivity of the liquid Conductivity depends on liquid type concentration temperature Change of...

Страница 27: ...y Disconnect signal wires at flowhead terminals 2 3 Do not disconnect cable shield terminals 1 GND Make sure that open cable ends are clean dry and do not contact each other or any flow meter part Rep...

Страница 28: ...that open cable ends are clean dry and do not contact each other or any flow meter part Repeat verification Does MagCheck indicate insulation 10 MOhm Now NOT OK Field current cable defective low insul...

Страница 29: ...ilable for MagCheck Spare part BaaN ordering number MagCheck in suitcase complete replacement XN00121100 Power supply adapter 100 230V to 15 V DC with plug adapter for different plug standards XN00121...

Страница 30: ...mber even years after first evaluation use the Excel spreadsheet MagCheck_Verification_Listing XLS Take some print outs with you whenever you leave for verifications and document local situation and f...

Страница 31: ...bration test Data and readings in menu 1 2 to 1 7 during manual calibration test may be noted in a print out of spreadsheet MagCheck_Manual _Test_Report XLS These data may be easily transferred into t...

Страница 32: ...isburg TEL 49 0 203 301 0 FAX 49 0 203 301 389 e mail krohne krohne de India KROHNE Marshall Ltd A 34 35 M I D C Industrial Area H Block Pimpri Poona 411018 TEL 91 0 20 744 20 20 FAX 91 0 20 744 20 40...

Отзывы: