KROHNE MagCheck Скачать руководство пользователя страница 14

 
 
 
 

14  

 

MagCheck

 

 

 
 
 

 

3.2.2  Explanation of the automatic verification run (verified data) 

 

Converter data 

Data indicated and stored  

Measurement of the field current  

Calibrated and actual value [mA] 

Measurement of field frequency 

Act frequency [Hz]  

ADC linearity and accuracy test at 25% of full scale 

1)

 

ADC linearity and accuracy test at 50% of full scale 

1)

 

ADC linearity and accuracy test at 75% of full scale 

1)

 

ADC linearity and accuracy test at 100 % of full scale 

1)

 

Output of input amplifier and 
analogue to digital converter: True 
and act. value [% of full scale] 

Testing of the 1st calibration point P04 for current output 

2)

 

Testing of the 2nd calibration point P20 for current output 

2)

 

Calibrated and actual value [mA] 

Testing of the 1st pulse output with f = 500 Hz  

3)

 

True and actual value [Hz] for 
f100% = 1000Hz 

If flowhead is tested as well  

 

Measurement of field coil resistance 

Coil resistance (Ohms) 

Measurement of resistance of electrode 1 

4)

 

Measurement of resistance of electrode 2 

4)

 

Measurement of isolation resistance of field coils 

Resistance values are measured. 
Display of subsequent voltage 
levels and resistance values  

 
Following the measurement, the measured data are stored in the MagCheck.  
 
1)  ADC test verifies accuracy of input amplifier and analogue to digital conversion of flow signal  

During this test the GK of the converter is temporarily set to factory test conditions GK = 2.500. 
At end of verification the GK will be reprogrammed automatically by MagCheck to original data. 

2)  mA output tests verify compliance of mA output with factory calibration settings  

Time constant T settings via IMoCom is reduced by MagCheck to allow for sufficiently fast 
settling of mA output and pulse output. At the end of the verification, the time constant will be 
reprogrammed automatically by MagCheck to original data. 

3)  For test of frequency / pulse output setting of this output will be temporarily programmed by 

MagCheck to 1000 Hz at full scale flow. At end of verification frequency / pulse output will be 
reprogrammed automatically by MagCheck to original data. 

4)  If mode "Empty pipe" has been selected under MagCheck  Function 1.1 electrode insulation 

data are measured, displayed and stored. 

 

In case of power interruption during flow meter verification e.g. by interruption of wiring, field current 
supply etc. the setting of GK, pulse output and time constant must be checked and eventually 
reprogrammed manually. Such an interruption is indicated by MagCheck after its restart with the 
warning message: Last meter not re-programmed. 
 
 
 
 

3.3 

MagCheck as manual flow signal calibrator 

 

3.3.1  Menu for manual calibration with IMoCom connection to IMoCom converters 

MagCheck IMoCom connector must be connected to the IMoCom connector of the converter.  
Note down all data settings and readings in a paper print of  the Excel spreadsheet 
“MagCheck_Verification_Listing.XLS” (see Appendix 1). After transfer of these data into this file, 
error calculations and a protocol will be created automatically. Main menus must be selected 
manually. 
 

 
 
 
 
 
 

Содержание MagCheck

Страница 1: ......

Страница 2: ...lead to loss of warranty In addition the general conditions of sale are applicable and are the basis for the purchase contract If the instrument needs to be returned to KROHNE please note the informat...

Страница 3: ...ual flow signal calibrator 14 3 4 MagCheck Menu functions for manual calibration test 17 4 Using MagCheck with a PC 21 4 1 System specifications 21 4 2 Installation PC software 21 4 3 Settings of MagC...

Страница 4: ...ersions Verification of the conventional electromagnetic flowmeter part of TIDALFLUX IFM 4110 PF is possible after disabling the level input to IFC 110 PF This will allow for a statement of flow veloc...

Страница 5: ...librator mode no data will be stored in MagCheck The measured indicated data must be noted and evaluated manually see Appendix 1 With non IMoCom converters In the flow signal calibrator mode with thes...

Страница 6: ...ow during the verification the meter Before changing any connections act according all process related safety and information rules existing in your organisation Disable all alarms and take all contro...

Страница 7: ...plied with MagCheck is only needed during download of verification data into a PC 2 3 Connection diagrams For use of MagCheck as precision flow signal calibrator for detailed manual verification of co...

Страница 8: ...8 MagCheck Connection MagCheck IFC 090 Connection MagCheck IFC 110 F...

Страница 9: ...EEPROM data for their validity After it will signal OK in case of an error it will indicate ERROR for approximately 5 seconds 2 5 1 With IMoCom converters IMoCom interface connected Next step is autom...

Страница 10: ...inus Correct field current is switching from positive to negative current and vice versa Plus Positive DC or pulsating DC defective converter negative half wave missing Minus Negative DC or pulsating...

Страница 11: ...hanging to submenus 3 1 l a n g u a g e To submenu l a n g u a g e E n g l i s h G B I M O C o m B u s F r e n c h F f i e l d c u r r e n t Back to main menu Ge r m a n D 2 6 3 Initial settings of Ma...

Страница 12: ...ro setting 1 4 Simulation Generation of flow signals adjustable in 0 1 steps of full scale value 1 5 Curr output Display of true value IMoCom converters only and actual value 1 6 Pulse output Display...

Страница 13: ...r at download into PC You may use up to 16 characters Next screen measurement Test conditions must be defined pipe full filled with liquid Electrode resistance will be tested via liquid With empty pip...

Страница 14: ...the GK will be reprogrammed automatically by MagCheck to original data 2 mA output tests verify compliance of mA output with factory calibration settings Time constant T settings via IMoCom is reduced...

Страница 15: ...e act value mA 1 6 Pulse output 1 6 Pulse output Shows pulse output at flow as entered under 1 4 True value Hz Actual value Hz Note true value Hz Note act value Hz 1 7 Field current 1 7 Field current...

Страница 16: ...l T900 000 1 3 Zero set 1 3 Zero set Allows MagCheck zero adjustment Physical unit as set under 1 2 Default 000 000 For standard tests use default For precision tests follow 3 4 2 1 4 Simulation 1 4 S...

Страница 17: ...ximum and minimum full scale values are subject to nominal diameter DN and the flow velocity v These limits are given in KROHNE documentation e g Installation and Operating Instructions of related con...

Страница 18: ...en 999 9 and 999 9 of the full scale as set or read under Function 1 2 Full Scale Value For normal tests setting shall be within 100 0 If the maximum signal level of MagCheck is exceeded the warning e...

Страница 19: ...hed Hardware Information and Status Information acc to menu 2 2 of IFC 010 020 090 110 are shown Display of Software I D no and status message of the connected module In this menu an 8 position softwa...

Страница 20: ...e supplied with MagCheck must be connected to the MagCheck connector the 9 pole connector to RS232 COM port of the PC COM port On the PC the MagCheck PC program must be active During transfer of the d...

Страница 21: ...for a password To work without password leave the field blank and click ok There is no way to recover a lost password After installation and clicking on the MagCheck icon on the desktop the main scre...

Страница 22: ...indicate Ready for data transfer Click on MagCheck PC main menu button Read flowmeter verification files from MagCheck MagCheck PC program will ask Number of new flowmeter verification files X e g 1...

Страница 23: ...ox allows for entering an appropriate file name e g MagCh_Date_All_UserComp MCK The list below shows all verification data files stored in file MagCheck MCK in the MagCheck data base Deselect files wh...

Страница 24: ...nd all settings as read by MagCheck Pull down lists will allow selection of a sequence of meter verifications in the meter history As an example the following view shows meter FT 1 parameter ADC 100 f...

Страница 25: ...selection menu This menu allows for selection of the device the parameter sequence to be printed print preview on the monitor printing as Adobe Acrobat Reader PDF file which may be stored or sent by...

Страница 26: ...ged 5 3 Electrode resistance with full pipe In general Electrode resistance with full pipe depends on conductivity of the liquid Conductivity depends on liquid type concentration temperature Change of...

Страница 27: ...y Disconnect signal wires at flowhead terminals 2 3 Do not disconnect cable shield terminals 1 GND Make sure that open cable ends are clean dry and do not contact each other or any flow meter part Rep...

Страница 28: ...that open cable ends are clean dry and do not contact each other or any flow meter part Repeat verification Does MagCheck indicate insulation 10 MOhm Now NOT OK Field current cable defective low insul...

Страница 29: ...ilable for MagCheck Spare part BaaN ordering number MagCheck in suitcase complete replacement XN00121100 Power supply adapter 100 230V to 15 V DC with plug adapter for different plug standards XN00121...

Страница 30: ...mber even years after first evaluation use the Excel spreadsheet MagCheck_Verification_Listing XLS Take some print outs with you whenever you leave for verifications and document local situation and f...

Страница 31: ...bration test Data and readings in menu 1 2 to 1 7 during manual calibration test may be noted in a print out of spreadsheet MagCheck_Manual _Test_Report XLS These data may be easily transferred into t...

Страница 32: ...isburg TEL 49 0 203 301 0 FAX 49 0 203 301 389 e mail krohne krohne de India KROHNE Marshall Ltd A 34 35 M I D C Industrial Area H Block Pimpri Poona 411018 TEL 91 0 20 744 20 20 FAX 91 0 20 744 20 40...

Отзывы: