7.2 Error
messages
Fct. Text
Description
and settings
4.0
PARAM.
ERROR
Main menu 4.0 Parameter error
4.1.0
FLOW
VELOC.
FLOW VELOCITY “v” incorrect:
Ensure condition
0.5 m/s
≤
v
≤
60 m/s
or 1.5 ft/s
≤
v
≤
180 ft/s is met!
4.1.1
FULL SCALE Full-scale range for flowrate Q
100%
,
see Fct. 3.1.1
4.1.2
METER SIZE Meter size, see Fct. 3.1.5
4.2.0
F/R FLOW
FULL-SCALE RANGE(S) for
forward/reverse flow incorrect:
Ensure condition F
≥
R is met!
4.2.1
FULL SCALE Full-scale range for forward
flowrate Q
F100%
, see Fct. 3.1.1
4.2.2
REV. SCALE Different range for reverse flow
required? See Fct. 3.1.2
4.2.3 REV.
VALUE
Full-scale range for reverse flow
Q
R100%
, see Fct. 3.1.3
4.3.0
I RANGE
CURRENT OUTPUT I RANGE
incorrect:
Ensure condition
I
100%
−
I
0%
≥
4 mA is met!
4.3.1
I 0 PCT
Current for 0% flow (I
0%
), see Fct.
3.3.3
4.3.2
I 100 PCT
Current for 100% flow (I
100%
), see
Fct. 3.3.4
4.4.0
I MAXIMUM
CURRENT LIMITATION
incorrect:
Ensure condition
I
max
≥
I
100%
is met!
4.4.1
I 100 PCT
Current for 100% flow (I
100%
), see
Fct. 3.3.4
4.4.2
I MAX mA
Setting of max. output current
(I
max
),
see Fct. 3.3.5
4.5.0
LFC. I RANG. LOW-FLOW CUTOFF RANGE I
incorrect:
Ensure condition
cutoff “off”.
Cutoff “on”
≥
1% is met!
4.5.1 L.F.
CUTOFF
I
Low-flow cutoff (SMU) for current
output required? See Fct. 3.3.7
4.5.2
CUTOFF ON Cutoff “on” value SMU-I, see Fct.
3.3.8
4.5.3 CUTOFF
OFF
Cutoff “off” value SMU-I, see Fct.
3.3.9
4.6.0
LFC. F
RANG.
LOW-FLOW CUTOFF RANGE F
incorrect:
Ensure condition cutoff “off”.
Cutoff “on”
≥
1% is met!
4.6.1 L.F.
CUTOFF
F
Low-flow cutoff (SMU) for
frequency output required?
See Fct. 3.4.6
4.6.2
CUTOFF ON Cutoff “on” value SMU-F,
see Fct. 3.4.7
4.6.3 CUTOFF
OFF
Cutoff “off” value SMU-F,
see Fct. 3.4.8