48
To set the Time Constant :
Begin from measuring mode.
This filtering only applies to mass and volume flow readings and any outputs that use them. The
mass totaliser is independent of the time constant.
The standard range for the time constant is 0.5 to 20 seconds. For faster response times, a
range of 0.2 to 20 seconds is available as a factory set option.
Time constant characteristic
5.4
Programming the display for measurement values (Fct. 1.2. and 3.2.)
The following measurement functions can be displayed :
Key
Display
line 1
line 2
→→→
Fct. 1.1.(1)
ZERO SET
↑↑
Fct. 1.1.(3)
TIME. CONST.
→
(0)4.0
TIME C. S.
Edit the time constant in
seconds in range 0.5 to 20.
↵
Fct. 1.1.(3)
TIME.CONST
4x
↵
Fct. 1.2.1
CYCL.DISP
Fct. 1.2.2
STATUS MSG.
Fct. 1.2.3
MASS FLOW
Fct. 1.2.4
MASS TOTAL
Fct. 1.2.5
DENSITY
Fct. 1.2.6
TEMPERAT.
Fct. 1.2.7
VOLUME.FLOW
Fct. 1.2.8
VOL.TOTAL
Содержание CORIMASS MFM 4085 K/F
Страница 26: ...26 4 2 Krohne Operating Concept...
Страница 54: ...54 Current output characteristics...
Страница 101: ...101 10 4 Block diagram of Converter MFC 085...