16
On Titanium tube meters, it is, however, possible to exceed 150°C up to an absolute maximum
of 200°C, for short periods. The time is limited by the initial temperature and final temperature.
To assess the time allowable, refer to figures 11,12 and 13, which cover the 10G+, 100G+ and
300G+ meters. For the 800G+ and 1500G+ refer to Krohne Ltd.
This does not apply to the 3000G+. These meters must not exceed 130°C. Zirconium tube
meters must never exceed 100°C.
ALLOWABLE TIME AT HIGH TEMP .
10 G METER
0
10
20
30
40
50
60
70
80
90
100
110
120
130
120
130
140
150
160
170
180
190
200
TEMP. C
TIME MIN.
Initial temp 30 C
Initial temp 70 C
Initial temp 120 C
ALLOWABLE TIME AT HIGH TEMP.
100G METER
0
10
20
30
40
50
60
70
80
90
100
110
120
120
130
140
150
160
170
180
190
200
TEMP. C
TIME MIN.
Initial temp 30 C
Initial temp 70 C
Initial temp 120
C