6.5 Error messages
Error Message
Description
Category
Remedy
NOT
LINEARIZED
Linearization
faulty or not
activated =
measuring
error
Errors
Activate linearization or carry it out again (HART
™
communication and linearization software are
required; the original calibration values must be
known), or send the device back to KROHNE for
linearization.
NEW LINEARI.
TABLE BAD
Faulty or non-
existent data in
the
linearisation
table =
measuring
error
Error
Check linearization or carry it out again (HART
™
communication and linearization software are
required; the original calibration values must be
known), or send the device back to KROHNE for a
check of the linearization.
LINEARIZATIO
UNDER
CONFIG
The device is in
linearization
mode =
measuring
error
Error
Complete the linearization and activate it (HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
UNIT SYSTEM
CONFLICT
The unit of the
linearisation
value is not
correct for the
select level
meter.
Error
Correct error, carry out linearization again if necessary
(HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
TOO FEW
ENTRIES
The
linearisation
table does not
have enough
support points.
Error
Carry out linearization at at least 5 points (HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
NOT
MONOTONIC
The sequence
of the
linearisation
values does not
rise
monotonously.
Error
Check linearization and/or carry it out again (HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
FIRST NOT 0 %
The first level
value of the
linearisation
table is not 0%
Error
Check linearization and/or carry it out again (HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
LAST NOT 100
%
The last level
value of the
linearisation
table is not
100%
Error
Check linearization and/or carry it out again (HART
™
communication and linearization software are
required), or send the device back to KROHNE for
linearization.
NO ZERO CAL
OF AO
The analog
output zero
point 4.00 mA
is not
calibrated. =
Possible
measurement
error in the
process control
system
Warning
Perform calibration using ammeter and menu item
3.10 or using standard HART
™
tools/process control
equipment and poss. external ammeter. Caution:
during the calibration, switch the measuring point to
manual control