Installation and operating instruction ALTOFLUX M900 and 3080 K 

 

 

Important information for installation:  PLEASE NOTE ! 

 

• 

Temperatures 

 

Refer to Section 11 “Limits” for operating pressure and vacuum load based on flange 
standards and type of tube liner. 

 
 

Ambient temperature 

Process temperature 

Compact system 3080 K 

-25 to +

60

 °C (-13 to +

140

 °F)  -25 to 

 +

60

 °C (-13 to 

 +

140

 °F) 

 

-25 to +

40

 °C (-13 to +

104

 °F)  -25 to 

>

 +

60

 °C (-13 to 

 +

194

 °F) 

ALTOFLUX M900 

-25 to +60 °C (-13 to +140 °F)  -25 to 

>

 +

60

 °C (-13 to 

>

 +

140

 °F) 

In storage 

-20 to +60 °C (-  4 to +140 °F), (Neoprene liner) 
-25 bis +60 °C (all other liners) 

 

avoid moisture and sunlight. 

Transport 

-  5 to +50 °C (-  4 to +140 °F), (Neoprene liner) 
-25 bis +60 °C (all other liners) 

 

avoid moisture and sunlight. 

 
 

• 

Location and position as required,

  

but electrode axis  

X – • – • – • – X

 

must be approximately horizontal in a 
horizontal pipe run. 
 

Y

  terminal box or converter housing 

 

 

 

• 

Measuring tube must be completetly filled at all times. 

 

• 

Direction of flow is arbitrary. 

Arrow on flowmeter can normally be ignored. For exceptions, 

refer to Section “Factory settings” in the installation and operating instructions for the signal 
converter. 

 

• 

Stud bolts and nuts: 

to fit, make sure there is sufficient room next to the pipe flanges. 

 

• 

Vibration:

 support the pipeline on both sides of the compact flowmeter. Level of vibration in 

conformity with IEC 068-2-34: below 2.2g for compact flowmeters in the frequency range of 
20-150 Hz with the IFC 090 K. 

 

• 

Do not expose to direct sunlight, 

fit a sunshade if necessary, not included with flowmeter,  

to be provided by customer. 

 

• 

Large meter sizes (

 DN 200 / 

 8”),

 use adapter pipes to allow axial shifting of the 

counterflanges and to facilitate installation. 

 

• 

Strong electromagnetic fields, 

avoid in vicinity of flowmeter 

 

• 

Straight inlet run minimum of 5 

×

 DN and outlet run minimum of 2 

×

 DN,

  

(DN = meter size), measured from the electrode axis.

 

Содержание ALTOFLUX 3080 K

Страница 1: ...packing of the primary head Storage and transport Pages 3 4 Installation in the pipeline Pages 4 5 7 8 Grounding Page 9 Primary head Compact flowmeters Variable area flowmeters Vortex flowmeters Flow...

Страница 2: ...10 300 3 8 12 Q100 0 1 3050 m 3 hr 0 37 13 440 US GPM ALTOFLUX 3080 K DN 10 300 3 8 12 Q100 0 1 3050 m 3 hr 0 37 13 440 US GPM This is equivalent to a flow velocity of 0 3 12 m s or 1 40 ft s Product...

Страница 3: ...grounding see Section 7 Grounding Certificate of calibration data Certificate of calibration data Grounding rings optional if ordered Grounding rings optional if ordered Installation instructions Inst...

Страница 4: ...minal box or converter housing Measuring tube must be completetly filled at all times Direction of flow is arbitrary Arrow on flowmeter can normally be ignored For exceptions refer to Section Factory...

Страница 5: ...ed either downstream of the flowmeter or upstream and downstream of the flowmeter 2 Suggestions for installation To avoid measuring errors due to gas air inclusion or to pipe running empty please obse...

Страница 6: ...verter by signal and field current cables ALTOFLUX 3080 K Compact flowmeter K IFC 090 K signal converter mounted direct on the primary head Versions for hazardous locations M900 and 3080 K are approve...

Страница 7: ...ipe run to avoid an excessive vacuum condition atthe meter Do not remove or damage liner which is formed around the flange edges Gaskets Use gaskets suitable for the application and appropriate to the...

Страница 8: ...7 4 8 4 4 3 2 32 40 4 x M 16 37 26 8 19 13 0 11 2 580 4 x 1 2 13 9 4 12 8 7 40 40 4 x M 16 43 31 1 25 18 1 2 580 4 x 5 8 24 17 4 23 16 6 50 40 4 x M 16 55 39 8 31 22 4 3 360 4 x 5 8 43 31 1 39 28 2 65...

Страница 9: ...cur in the pipeline system An FE functional ground must simultaneously take over the function of the protective conductor combined protective functional ground Refer to appropriate national codes for...

Страница 10: ...tion and operating instructions for the signal converter 4 Specific calibration data are defined during factory calibration for each primary head which are indicated on the instrument nameplate This i...

Страница 11: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 11 Notes...

Страница 12: ...12 Installation and operating instruction ALTOFLUX M900 and 3080 K 9 Technical data...

Страница 13: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 13...

Страница 14: ...14 Installation and operating instruction ALTOFLUX M900 and 3080 K 10 Dimensions and weights...

Страница 15: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 15...

Страница 16: ...16 Installation and operating instruction ALTOFLUX M900 and 3080 K...

Страница 17: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 17 11 Limits...

Страница 18: ...18 Installation and operating instruction ALTOFLUX M900 and 3080 K Notes...

Страница 19: ...lowmeter is safe to handle If the flowmeter has been operated with toxic caustic flammable or water endangering liquids you are kindly requested to check and ensure if necessary by rinsing or neutrali...

Отзывы: