Installation and operating instruction ALTOFLUX M900 and 3080 K 

 

7 Grounding 

• 

All flowmeters must be properly grounded to avoid personnel shock hazard. 

• 

The ground conductor should not transmit any interference voltages, therefore do not ground any 
other electrical devices together with this conductor. 

 

ALTOFLUX  M900

  separate primary head with terminal box 

 

• 

An 

FE functional ground

 must always be connected. 

• 

Signal converter with field power supply > 125 mA / 60 V

 a 

PE protective conductor

 must be 

connected to the primary head, because of the higher field current from the signal converter. See 
grounding diagrams below. 

 

ALTOFLUX  3080 K

  compact systems 

 

Supply power > 50 V AC 

• 

Grounding is via the 

PE protective ground conductor

 incorporated in the power supply cable,  

see also Section “Connection to power” in the installation and operating instructions for the signal 
converter. 

• 

EXCEPTION:

 

Do not connect up the PE protective ground conductor in the terminal box

  

if e.g. compact units are operated in the proximity of electric furnaces, electrolysis plants, etc.,  
and large potential differences occur in the pipeline system. An FE functional ground must 
simultaneously take over the function of the protective conductor (combined protective/functional 
ground). Refer to appropriate national codes for specific requirements for this type of installation, 
which may require the addition of a ground fault detection circuit interrupter. 

 

Power supply 24 V AC or DC 

• 

Protective separation (PELV) must be ensured (VDE 0100 / VDE 0106 or IEC 364 / IEC 536 or 
equivalent national regulations). 

• 

An 

FE functional ground conductor

 must be connected for measurement reasons. 

 

Grounding diagrams 

 

Metal pipelines, 

Metal pipelines, 

not internally coated 

with or without internal coating, 

 

and plastic pipelines 

grounding without grounding rings 

grounding with grounding rings 

 

 

 

D1, D2, D3 

Gaskets, not included with supply, to be provided by customer. 

Grounding rings (option) 

Flowmeter flanges 

FE 

Functional ground, wire

 ≥

 4 mm

2

 Cu (10 AWG), not included with flowmeter, to be provided by customer 

PE 

Protective conductor required if the ALTOFLUX M900 is operated with a signal converter that supplies a field current of 

> 125 mA / > 60 V

. Wire 

 4 mm

2

 Cu (10 AWG), not included with flowmeter, to be provided by customer. 

Pipeline 

RF 

Pipe flanges 

V1, V2 

Interconnecting wires, included with flowmeter 

Terminal box or signal converter 

Содержание ALTOFLUX 3080 K

Страница 1: ...packing of the primary head Storage and transport Pages 3 4 Installation in the pipeline Pages 4 5 7 8 Grounding Page 9 Primary head Compact flowmeters Variable area flowmeters Vortex flowmeters Flow...

Страница 2: ...10 300 3 8 12 Q100 0 1 3050 m 3 hr 0 37 13 440 US GPM ALTOFLUX 3080 K DN 10 300 3 8 12 Q100 0 1 3050 m 3 hr 0 37 13 440 US GPM This is equivalent to a flow velocity of 0 3 12 m s or 1 40 ft s Product...

Страница 3: ...grounding see Section 7 Grounding Certificate of calibration data Certificate of calibration data Grounding rings optional if ordered Grounding rings optional if ordered Installation instructions Inst...

Страница 4: ...minal box or converter housing Measuring tube must be completetly filled at all times Direction of flow is arbitrary Arrow on flowmeter can normally be ignored For exceptions refer to Section Factory...

Страница 5: ...ed either downstream of the flowmeter or upstream and downstream of the flowmeter 2 Suggestions for installation To avoid measuring errors due to gas air inclusion or to pipe running empty please obse...

Страница 6: ...verter by signal and field current cables ALTOFLUX 3080 K Compact flowmeter K IFC 090 K signal converter mounted direct on the primary head Versions for hazardous locations M900 and 3080 K are approve...

Страница 7: ...ipe run to avoid an excessive vacuum condition atthe meter Do not remove or damage liner which is formed around the flange edges Gaskets Use gaskets suitable for the application and appropriate to the...

Страница 8: ...7 4 8 4 4 3 2 32 40 4 x M 16 37 26 8 19 13 0 11 2 580 4 x 1 2 13 9 4 12 8 7 40 40 4 x M 16 43 31 1 25 18 1 2 580 4 x 5 8 24 17 4 23 16 6 50 40 4 x M 16 55 39 8 31 22 4 3 360 4 x 5 8 43 31 1 39 28 2 65...

Страница 9: ...cur in the pipeline system An FE functional ground must simultaneously take over the function of the protective conductor combined protective functional ground Refer to appropriate national codes for...

Страница 10: ...tion and operating instructions for the signal converter 4 Specific calibration data are defined during factory calibration for each primary head which are indicated on the instrument nameplate This i...

Страница 11: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 11 Notes...

Страница 12: ...12 Installation and operating instruction ALTOFLUX M900 and 3080 K 9 Technical data...

Страница 13: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 13...

Страница 14: ...14 Installation and operating instruction ALTOFLUX M900 and 3080 K 10 Dimensions and weights...

Страница 15: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 15...

Страница 16: ...16 Installation and operating instruction ALTOFLUX M900 and 3080 K...

Страница 17: ...Installation and operating instruction ALTOFLUX M900 and 3080 K 17 11 Limits...

Страница 18: ...18 Installation and operating instruction ALTOFLUX M900 and 3080 K Notes...

Страница 19: ...lowmeter is safe to handle If the flowmeter has been operated with toxic caustic flammable or water endangering liquids you are kindly requested to check and ensure if necessary by rinsing or neutrali...

Отзывы: