12.3 Maintenance plan
Maintenance and cleaning work must be performed at regular intervals as specified in the
table below.
Interval
Cleaning work
daily
The
humidifier
must be cleaned and disinfected every day.
daily
(when used in homes or
for nursing care)
Clean the
nasal cannula
.
every 14 days
earlier if necessary
The
aeroplus m
must be cleaned with a damp cloth and
disinfected after.
every 4 weeks
Replace the
coarse dust filter
every 4 weeks
(when used in homes or
for nursing care)
Replace the
nasal cannula
.
annually / after 5000
operating hours
Replace the
air intake filter
NOTE!
With extremely dirty ambient air the filter must be
changed earlier.
with changing patients
The
aeroplus m
must be cleaned with a damp cloth and
disinfected after.
with changing patients
Replace the
humidifier
.
with changing patients
Replace the
nasal cannula
.
WARNING! Health risk!
In order to avoid cross infection, each user of the
aeroplus m should wear his/her own nasal cannula.
after infections
After an infection a new
nasal cannula
should be used to
prevent re-infection.
after service work
The
aeroplus m
must be cleaned with a damp cloth and
disinfected after.