aeroplus m
Messages and Alarms
41
Display/Message
Description
High battery
temperature
Temperature of the battery is too high for the charging
process. Charging function is interrupted.
Counter measures:
Take the aeroplus m into a lower ambient temperature.
NOTE!
The charging process will automatically continue at a battery
temperature of < 42° C.
11.3 Alarms
Following low-priority alarms can be triggered:
Alarm category/
Display
Description
Oxygen
The product gas does not have the specified oxygen
concentration.
The oxygen concentrator aeroplus m
is equipped with an innovative
multifunctional sensor that monitors constantly the oxygen
concentration of the production process.
In case of a deviation from the technical specifications, the oxygen
alarm is triggered.
Counter measures
with constant alarms, consult technical service.
Pressure
The pressure in the oxygen reservoir is too low.
Counter measures:
Re-boot the unit. If alarm still continues, contact the technical
service.