aeroplus m
Menu and advanced settings
32
10.5 Lower limit for the therapy setting
In order to limit the therapy range, a lower (minimum oxygen volume, O2-Min) and an upper
limit (maximum oxygen volume, O2-Max) can be stored. These limits ensure the operation
only between the applied limits.
Display/Message
Description
To set the lower limit, select menu item
O2-Min.
in the main menu;
the selection is inverted with white back ground color.
NOTE!
Already in the main menu, the current setting can be controlled
(here: setting 0.5). It is not necessary to call the submenu.
Confirm selected menu item with the menu button
.
To increase the setting press the arrow-up-button
.
To decrease the setting press the arrow-down-button
.
To leave this menu, press the menu button
(=
return
)
.
NOTE!
If the lower limit has the same value as the upper limit, this will
freeze the therapy setting.
NOTE!
The lower limit can only be set between 0.5 and the upper
therapy limit. If applicable, the upper limit must be modified
first; not till then the lower limit can be adjusted as desired.