EN
DMM3000
Range
OL Reading
Resolution
Accuracy
600.0mV
660.0mV
0.1mV
±(1.0% + 5D)
6.000V
6.600V
0.001V
±(1.0% + 3D)
60.00V
66.00V
0.01V
600.0V
660.0V
0.1V
1000V
1100V
1V
LCD displays 0 counts when the reading < 10 counts.
Input Impedance:
0ȍOHVVWKDQS)
Frequency Response:
45~500Hz(Sine Wave)
Overload Protection:
AC/DC 1000V
5. AC Voltage
Range
OL Reading
Resolution
Accuracy
600.0V
660.0V
0.1V
±(2.0% + 3D)
1000V
1100V
1V
LCD displays 0 counts when the reading < 10 counts.
Input Impedance :
OHVVWKDQNȍ
Frequency Response :
45 ~ 500Hz (Sine Wave)
Overload Protection :
AC/DC 1000V
6. Auto Voltage
Range
OL Reading
Resolution
Accuracy
6.000A
6.600A
0.001A
±(1.0% + 3D)
10.00A
20.00A
0.01A
Maximum measurement time :
>5A for max.3 minutes with at least 20 minutes rest time.
>10A for max.30 seconds with at least 10 minutes rest time.
Overload Protection :
AC/DC 11A
7. DC Current
15