5
Power
On/Off(press and hold for 5 seconds)
On/Off(press and
hold for 5
secondseconds)
On/Off(press and hold for 5
seconds)
Print instant gas concentration data
Invalid
Invalid
CAUTION:
1. All operations are operate by a short press of the button unless it is special noticed
2. GT-1000 series multifunctional gas and dust detector could be work normally only in condition of the pump was switched on
②
Gas Detector Processing Operation under Out-range Status
Users should avoid to have sensor impacted by the gas with a pressure value greater than the maximum of the detector, which might
affect the service life and precision of the detector, or even directly damage detector.
When a user accidentally makes out-range operation, he should evacuate the instrument out of the detecting site and place it in the
clean air for more than half an hour. During the time, user should observe whether the density value of the instrument is keeping
decreasing or not. If it can straightly go down to normal value, then he can continue to use it after the zero calibration of instrument.
While the instrument after the out-range operation and user has placed it in clean air for hours, the density value remains high, then
it should be sent back to the manufacturer or agent for maintenance, be ready to replace the sensor.
Special Note: Detector damage resulted from out-range operation is not within the warranty.