Konica Minolta Ultra KDR Скачать руководство пользователя страница 9

 

Ultra KDR Panel Configuration and Calibration Guide v1.1

 

Ultra

 

9

 

 

 

 

 

 

           

Customer Support 

1.800.366.5343 – 

[email protected]

 

VZ0099HTG201606-1.1 

 

 

8.  At the 

Please input panel Id: 

 prompt, 

type the Panel Serial Number listed on the 
back of the panel. 
 

 

Figure 18 Aero SDK Library Sample Program 
command line 

9.  Press Enter. 

10. At the 

XG-IFBox No.

 -

 prompt, type “1” 

and press Enter. 

11. At the 

Please input Number Of Trials [1-

32] 

 prompt, type “8” (Recommended) 

and press Enter. 

12. After configuration is complete, the 

application will display the message, 
Exposure Waiting. 
 

 

Figure 19 

13. In Tech Service, adjust techniques to 80 

KvP, 10ms, and 100 mA. 

14. When ready, take your first exposure. 

15. If the “Recommended Technique” is low or 

high. The application will prompts you to 
adjust your technique by a percentage (Ex: 
Signal too Low, Adjust by 25%). 

16. If technique meets requirements continue 

with next exposure when the Exposure 
Waiting message is displayed. 
 

 

Figure 20 

17. Continue, repeating exposure until you have 

reached number of trials that was defined. 

18. Once the last exposure is complete, the 

following message is displayed. 

 

Figure 21 

19. Close the sample application. 

20. Exit Tech Service. 

1.4.3  Verify calibration 

It is important to test exposures using a test patient 
to verify calibration. For information on how to take 
exposures, see the “Ultra DR User Quick Guide”. 

 

Содержание Ultra KDR

Страница 1: ...Ultra KDR Configuration and Calibration Guide...

Страница 2: ...el Registration Configuration and Calibration 3 1 1 Before you begin 3 1 2 Configuring the KDR panel 4 1 3 Panel Registration 4 1 3 1 Before you begin 4 1 3 2 Panel registration 5 1 3 3 Panel configur...

Страница 3: ...n Manual A9KXBB01EN01_160316 Before starting panel setup verify the network settings for the onboard card The following configuration applies to Ultra 4 1 0 x or later with the KDR Panel and driver 10...

Страница 4: ...cessing Default value 0 X Ray Synchronization Integrated XG IF Box ID 1 Exposure Time 1 0 Figure 3 Device configuration 1 3 Panel Registration 1 3 1 Before you begin You will need to power on the KDR...

Страница 5: ...3 On the Device Configuration tab under Device Identity section enter the Panel serial number Figure 6 You can find this number located on the back of the panel Figure 6 Device configuration 4 Click...

Страница 6: ...n the Device Identity list form in the IF cable packaging Figure 9 item 2 8 Click OK Figure 9 item 3 The KONICA MINOLTA AeroSDK Setting Tool main screen will become active 9 On the KONICA MINOLTA Aero...

Страница 7: ...ng Figure 13 Figure 13 Konica Minolta KDR CD 1 3 3 Panel configuration 1 Insert the installation CD The Aero Installer window is displayed Figure 14 Select the Browse To option Highlighted in yellow a...

Страница 8: ...to the off position The panel will power off after 60 seconds Switching back on at any time after the panel is turned off only requires the 30 second start up and initialization as instructed in Figu...

Страница 9: ...h Service adjust techniques to 80 KvP 10ms and 100 mA 14 When ready take your first exposure 15 If the Recommended Technique is low or high The application will prompts you to adjust your technique by...

Отзывы: