Ultra KDR Panel Configuration and Calibration Guide v1.1

 

Ultra

 

4

 

 

 

 

 

 

           

Customer Support 

1.800.366.5343 – 

[email protected]

 

VZ0099HTG201606-1.1 

 

 

1.2  Configuring the KDR panel 

1.  In Ultra, select the Device Configuration tab. 

2.  Under the Configuration Pool section, select 

Default

, then select 

KDR

 (Figure 3). 

3.  Under the device settings, use the following 

configuration values: 

Field 

Value 

Maven 

Aero, 32 Bit, #1 

Exposure Index 

Determined during 
installation 
(Default is 1000) 

Edge Crop 

Default value, 0 

AEC Display 

Default value, 0 

Global Orientation 

Default value, 0 

Image Processing 

Default value, 0 

X-Ray 
Synchronization 

Integrated 

XG-IF Box ID 

Exposure Time 

1.0 

 

 

Figure 3 Device configuration 

1.3  Panel Registration  

1.3.1  Before you begin 

You will need to power on the KDR panel and have 
the device identity list available (Figure 4), which is 
provided in the IF cable box. The device identity list 
includes information such as the serial number and 
Sub device of the IF cable. 

For detailed device information refer to KDR System Operation Manual 
(A9KXBB01EN01_160316) Chapter 2.

    

 

Figure 4 Device identity list 

Also on the front of the power box (

XG-IFBox

.), flip 

the switch to the on position as shown in (Figure 5)  

 

Figure 5 XG-IFBox 

After switch is flipped into the on position, please allow 
up to 30 seconds for the panel to start up and initialize. 

Содержание Ultra KDR

Страница 1: ...Ultra KDR Configuration and Calibration Guide...

Страница 2: ...el Registration Configuration and Calibration 3 1 1 Before you begin 3 1 2 Configuring the KDR panel 4 1 3 Panel Registration 4 1 3 1 Before you begin 4 1 3 2 Panel registration 5 1 3 3 Panel configur...

Страница 3: ...n Manual A9KXBB01EN01_160316 Before starting panel setup verify the network settings for the onboard card The following configuration applies to Ultra 4 1 0 x or later with the KDR Panel and driver 10...

Страница 4: ...cessing Default value 0 X Ray Synchronization Integrated XG IF Box ID 1 Exposure Time 1 0 Figure 3 Device configuration 1 3 Panel Registration 1 3 1 Before you begin You will need to power on the KDR...

Страница 5: ...3 On the Device Configuration tab under Device Identity section enter the Panel serial number Figure 6 You can find this number located on the back of the panel Figure 6 Device configuration 4 Click...

Страница 6: ...n the Device Identity list form in the IF cable packaging Figure 9 item 2 8 Click OK Figure 9 item 3 The KONICA MINOLTA AeroSDK Setting Tool main screen will become active 9 On the KONICA MINOLTA Aero...

Страница 7: ...ng Figure 13 Figure 13 Konica Minolta KDR CD 1 3 3 Panel configuration 1 Insert the installation CD The Aero Installer window is displayed Figure 14 Select the Browse To option Highlighted in yellow a...

Страница 8: ...to the off position The panel will power off after 60 seconds Switching back on at any time after the panel is turned off only requires the 30 second start up and initialization as instructed in Figu...

Страница 9: ...h Service adjust techniques to 80 KvP 10ms and 100 mA 14 When ready take your first exposure 15 If the Recommended Technique is low or high The application will prompts you to adjust your technique by...

Отзывы: