Konica Minolta Ultra KDR Скачать руководство пользователя страница 6

 

Ultra KDR Panel Configuration and Calibration Guide v1.1

 

Ultra

 

6

 

 

 

 

 

 

           

Customer Support 

1.800.366.5343 – 

[email protected]

 

VZ0099HTG201606-1.1 

 

 

5.  Click 

Add IF Cable

. (Figure 8) 

 

Figure 8 AeroSDK Setting Tool – IF Cable 
Registration 

The Konica Minolta AeroSDK Setting Tool: 
IF Cable Registration screen is displayed. 

6.  Verify that 

In my Room

 is selected (Figure 

9, item 1). 
 

 

Figure 9 AeroSDK Setting Tool – IF Cable 

Registration 

7.  In the 

IF Cable ID

 field, enter the 16 digit 

Sub device ID located on the Device Identity 
list form in the IF cable packaging (Figure 9, 
item 2). 

8.  Click 

OK

 (Figure 9, item 3). 

The KONICA MINOLTA AeroSDK Setting 
Tool main screen will become active. 

9.  On the KONICA MINOLTA AeroSDK 

Setting Tool main screen, click 

Add Panel

 

(Figure 9). 
 

 

Figure 10 AeroSDK Setting Tool - Add Panel 

The KONICA MINOLTA AeroSDK Setting 
Tool –Add Panel is displayed. 
 

 

Figure 11 AeroSDK Setting Tool - Add Panel 

10. Enter the Panel serial number located on 

the back of the panel. (Figure 11, item 1). 

11. Ensure that 

G1+/G2

 is selected for the 

Generation of panel you are using (Figure 
11, item 2). 

12. Keep the default 

Wired LAN, Wireless 

LAN IP and Subnet mask

 settings as listed 

in Figure 11, items 3, 4, 5, and 6. 

Содержание Ultra KDR

Страница 1: ...Ultra KDR Configuration and Calibration Guide...

Страница 2: ...el Registration Configuration and Calibration 3 1 1 Before you begin 3 1 2 Configuring the KDR panel 4 1 3 Panel Registration 4 1 3 1 Before you begin 4 1 3 2 Panel registration 5 1 3 3 Panel configur...

Страница 3: ...n Manual A9KXBB01EN01_160316 Before starting panel setup verify the network settings for the onboard card The following configuration applies to Ultra 4 1 0 x or later with the KDR Panel and driver 10...

Страница 4: ...cessing Default value 0 X Ray Synchronization Integrated XG IF Box ID 1 Exposure Time 1 0 Figure 3 Device configuration 1 3 Panel Registration 1 3 1 Before you begin You will need to power on the KDR...

Страница 5: ...3 On the Device Configuration tab under Device Identity section enter the Panel serial number Figure 6 You can find this number located on the back of the panel Figure 6 Device configuration 4 Click...

Страница 6: ...n the Device Identity list form in the IF cable packaging Figure 9 item 2 8 Click OK Figure 9 item 3 The KONICA MINOLTA AeroSDK Setting Tool main screen will become active 9 On the KONICA MINOLTA Aero...

Страница 7: ...ng Figure 13 Figure 13 Konica Minolta KDR CD 1 3 3 Panel configuration 1 Insert the installation CD The Aero Installer window is displayed Figure 14 Select the Browse To option Highlighted in yellow a...

Страница 8: ...to the off position The panel will power off after 60 seconds Switching back on at any time after the panel is turned off only requires the 30 second start up and initialization as instructed in Figu...

Страница 9: ...h Service adjust techniques to 80 KvP 10ms and 100 mA 14 When ready take your first exposure 15 If the Recommended Technique is low or high The application will prompts you to adjust your technique by...