How to Handle Film Flaws
How to Handle Film Flaws That Consistently Occur in the Same Position
If flaws consistently occur in the same position on printed film, it may be that foreign matter has been
deposited on the feed guides.
In such a case, clean the feed guides with a soft lint-free cloth dampened with ethanol. The feed guides
in the following 2 locations must be cleaned.
•
Inside the top right cover
•
Inside the lower right cover (when an expansion tray or mammography media expansion tray is at-
tached)
Wiping the feed guides too harshly may damage them, which in
turn would cause more film flaws. Therefore, wipe them carefully.
•
If film flaws affecting the interpretation of radiographs persist even after
cleaning the feed guides, contact Konica Minolta technical representatives.
Cleaning the feed guides inside the top right cover
Open the top right cover, and wipe the guide plate and roller with an ethanol-dampened cloth.
Cleaning the feed guides inside the lower right cover (when an expansion
tray or mammography media expansion tray is attached)
Open the lower right cover, and wipe the guide plate and roller with an ethanol-dampened cloth.
Содержание DRYPRO 873
Страница 1: ...EN 13...
Страница 2: ......
Страница 9: ...9 Chapter 1 Safety Warnings Precautions...
Страница 25: ...25 Chapter 2 Product Outline...
Страница 35: ...35 Chapter 3 Operation from Main Unit...
Страница 76: ...76...
Страница 77: ...77 Chapter 4 Troubleshooting...
Страница 100: ...100...
Страница 101: ...101 Chapter 5 Maintenance...
Страница 115: ...115 Appendix...
Страница 120: ...120...
Страница 121: ...121...
Страница 122: ...122...
Страница 123: ......