background image

5

Notes on Storage

• 

The CM-3700A-U should be stored at temperatures between 0°C and 40°C, and at a relative humidity of 
80% or less (35°C) without condensation. Do not store the instrument in areas subject to high tempera-
tures, high humidity, sudden changes in temperature, or where freezing or condensation may occur, be-
cause these circumstances may cause a breakdown. It is recommended to store the CM-3700A-U with a 
drying agent at a temperature around 20°C.

• 

Do not leave the CM-3700A-U inside a car such as in the trunk. Otherwise, the temperature and/or humid-
ity may exceed the allowable range for storage during midsummer or midwinter, resulting in a breakdown.

• 

Keep the packing materials used for shipment and use them to transport the CM-3700A-U. This protects 
the instrument from sudden changes in temperature, vibration, and shock.

• 

Do not store the CM-3700A-U in areas where dust, cigarette smoke or chemical gases are present. Doing 
so may cause deterioration in performance or a breakdown.

• 

Entry of dust into the specimen measuring port will hinder accurate measurement. When the instrument is 
not in use, you must attach the target mask and the sample holder, and cover the instrument with the sup-
plied Dust Cover to prevent the entry of dust into the integrating sphere.

• 

The White Calibration Plate may become discolored if left exposed to light. Therefore, make sure to close 
the cover when it is not in use so that the White Calibration Plate is not exposed to ambient light during 
storage.

• 

The Target Masks may discolor if they are left exposed to light. When they are not in use, keep them in a 
safe place to prevent exposure to light and to protect them from scratches and dust. And store them in the 
accessory case (CM-A156).

• 

Take care not to leave the CM-3700A-U for a long period of time with a target mask attached. The sample 
holder may stick to the target mask.

• 

Be sure to keep all packing materials (cardboard box, cushioning material, plastic bags, etc.). They can be 
used to protect the instrument during transportation to the service facility for maintenance (re-calibration 
etc.).

Notes on Cleaning

• 

If the CM-3700A-U becomes dirty, wipe it with a soft, clean dry cloth. Never use solvents such as thinner or 
benzene.

• 

If the White Calibration Plate becomes dirty, wipe it with a soft, clean dry cloth. If dirt is difficult 
to remove, wipe it off with a cloth dampened with commercially-available lens cleaning solution.                               
Then remove the solution with a cloth dampened with water, and leave the plate to dry.

• 

If the inner surface (white-coated surface) of the Target Masks, or the inside of the integrating sphere, get-
dirty, contact a Konica Minolta-authorized service facility.

• 

Should the CM-3700A-U break down, do not try to disassemble and repair it by yourself. Contact a Konica 
Minolta-authorized service facility.

Disposal Method

• 

Make sure that the CM-3700A-U and its accessories and packing materials are either disposed of or re-
cycled correctly in accordance with local laws and regulations.

Содержание CM-3700A-U

Страница 1: ...SPECTROPHOTOMETER CM 3700A U INSTRUCTION MANUAL Before using this instrument please read this manual...

Страница 2: ...le the instrument Denotes an instruction Be sure to disconnect the AC adapter from the AC outlet This symbol indicates A C This symbol indicates D C This symbol indicates class II protection against e...

Страница 3: ...t is damaged or the AC adapter is damaged or if smoke or odd smells occur Doing so may cause a fire In such situations turn the power OFF immediately disconnect the AC adapter plug from the AC outlet...

Страница 4: ...tion during shipment to Konica Minolta for maintenance Should they be lost or damaged contact the nearest Konica Minolta authorized service facility Protective cap for the specimen measuring port inte...

Страница 5: ...mage the instru ment Measurement Make sure no dirt or dust get into the specimen measuring port When using the instrument for long periods of time the measurement value may change depending on changes...

Страница 6: ...uld be taken to prevent them getting dirty or exposing them to impacts When the CM 3700A U is not in use be sure to attach a target mask to the measuring port to prevent entry of foreign matter The CM...

Страница 7: ...it is not in use so that the White Calibration Plate is not exposed to ambient light during storage The Target Masks may discolor if they are left exposed to light When they are not in use keep them i...

Страница 8: ...calibration data 17 Performing calibration 18 Attaching the Zero Calibration Box 19 Notes on Use of Zero Calibration Box 19 Attaching the White Calibration Plate 20 Notes on Use of White Calibration...

Страница 9: ...7 Using the CM 3700A U...

Страница 10: ...to the specimen Measurement and illumination aperture area at specimen surface areas when each target mask is attached are as follows CM A91 SAV 1 3 mm 5 7 mm CM A95 USAV 1 3 mm 3 5 mm Zero Calibratio...

Страница 11: ...ovides various functions e g data processing and file management and allows the user to operate the CM 3700A U using a personal computer Ver 2 6 or later Optional Accessories USB Cable 5 m IF A22 Used...

Страница 12: ...ing calibration data and data setting software Target Mask CM A91 SAV USB Cables IF A21 3m USB IF A22 5m Zero Calibration Box CM A155 Standard accessories Optional accessories CM S100w Professional Ed...

Страница 13: ...and OFF Sample holder Used to hold the specimen white calibration plate or zero calibration box Target mask Select a suitable target mask for 3 5 mm illumination USAV 5 7 mm illumination SAV accordin...

Страница 14: ...n the PC to be used to control the CM 3700A U Starting the software Start the software and set it for control of the CM 3700A U Turning the power ON Turn the power ON P 15 Attaching a calibration targ...

Страница 15: ...connect the USB cable connector check the shape of the receptacle connection terminal and insert the connector fully until it is secured Connecting a Personal Computer Operating Procedure In general a...

Страница 16: ...se electric shock Insert the power plug fully and securely Incomplete inserting may cause fire or electric shock Do not disassemble or modify the AC adapter Doing so may cause a fire or electric shock...

Страница 17: ...a fire In such situations turn the power OFF immediately disconnect the AC adapter from the AC outlet and contact the nearest Konica Minolta authorized service facility Turning Power ON and OFF Proce...

Страница 18: ...mo The target mask is held on by magnets 3 Align the target mask positioning holes of the target mask to be attached with the target mask positioning pins of the Spectrophotometer s specimen port base...

Страница 19: ...ration plate which has a reflectance of approximately 100 and for which spectral reflectance data are provided as white calibration data to set a high reflectance reference point for reflectance measu...

Страница 20: ...ed However to more accurately perform measurement it is recommended to perform zero calibration and white calibration every time measurement conditions are changed c c When the calibration has been pe...

Страница 21: ...Procedure 1 Pull the sample holder toward you and keep it open The sample holder will remain open when opened more than 70 degrees 2 Fit the projections of the zero calibration box into the grooves o...

Страница 22: ...The white calibration plate may become discolored if left exposed to light Therefore when not in use make sure that the lid is closed to prevent exposure to light Take care not to scratch touch or ma...

Страница 23: ...Specimen Reflectance measurement Before starting reflectance measurements attach the target mask to be used for measurements Before starting reflectance measurements set the measurement area and specu...

Страница 24: ...d Remove the two stopper attaching screws by turning them counterclockwise using a Phillips head screwdriver Keep the removed screws and stopper in a safe place When widely opening the measurement por...

Страница 25: ...g liquid and wipe Then wipe off the liquid with a cloth dampened with water and leave it to dry When cleaning take care not to scratch the zero calibration box or white calibration plate Target Mask U...

Страница 26: ...to emit light Xenon lamp dead Illumination circuit failure Sensor failure Retry measurement or perform recalibration If the message persists contact the nearest Konica Minolta authorized service facil...

Страница 27: ...ter and trap positions is in progress After turning the power on wait for ten and several seconds until the lens trap and UV filter positions are initialized Time for Periodic Calibration Please conta...

Страница 28: ...20 Was white calibration performed correctly Attach White Calibration Plate correctly and perform white calibration correctly 20 Was zero calibration performed correctly Attach Zero Calibration Plate...

Страница 29: ...27 Explanations...

Страница 30: ...m the illumination monitoring optical fiber is divided into each wavelength component and projected onto the sensor array sections which convert the light into proportional currents and output the cur...

Страница 31: ...rding to the specimen and your application The target mask surface which faces into the integrating sphere has a high reflectance white coating Since the reflectance of this coating affects measuremen...

Страница 32: ...30 Dimensions 170 6 170 6 37 27 247 6 75 5 91 5 197 371 271 2 496 478 71 8 91 5 mm...

Страница 33: ...ce Standard deviation within 0 05 Colorimetric values Standard deviation within E ab 0 005 When a white calibration plate is measured 30 times at 10 second intervals after white calibration Inter inst...

Страница 34: ...32 MEMO...

Страница 35: ...RESULTING FROM THE MISUSE MISHANDLING UNAUTHORIZED MODIFICATION ETC OF THIS PRODUCT OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS INTERRUPTION OF BUSI...

Страница 36: ...2015 2021 KONICA MINOLTA INC 9222 A4KX 21 CBGCDK...

Отзывы: