83

Measur

ement 

Section

7

. Adjust the white balance.

Normally, white balance is adjusted by adjusting the cutoff and drive voltages. However, in the procedure
below, the display is adjusted so that the white generated on the display matches the target color (W) stored
in memory.

The method is explained by taking the following cases
where the measured values are as follows compared to the
target color (W).

emission intensity of R: Higher by 20%

emission intensity of B: Lower by 10%

emission intensity of G: Higher by 10%

7-1. When analyzer mode (G-standard) is se-

lected

1

Adjust the luminance (or emission intensity of G)

so that the displayed value for G changes from

“110” to “100.0”.

The other value (R and B) may change somewhat if
the intensity is adjust.

2

Adjust the out put of R so that the displayed value

for R changes from “109.0” to “100.0”, and ad-

just the out put of B so that the displayed value

for B changes from “81.80” to “100.0”.

When all the values for R, B and G are changed to
“100.0”, adjustment of the white color of the display
to the target color (W) (i,e,the xyLv values are the
same) is complete.

7-2. When analyzer mode (R-standard) is se-

lected

1

Adjust the luminance (or emission intensity of R)

so that the displayed value for R changes from

“120” to “100”.

The other value (G and B) may change somewhat if
the intensity is adjust.

2

Adjust the out put of B so that the displayed value

for B changes from “75.00” to “100.0”, and adjust

the out put of G so that the displayed value for G

changes from “91.70” to “100.0”.

When all the values for R, B and G are changed to
“100.0”, adjustment of the white color of the display
to the target color (W) (i,e,the xyLv values are the
same) is complete.

* The RGB values given in the above example are based on calculations, and may not correspond to the actual display.

R

B

G

R

B

G

G

120.0

90.00

110.0

R

B

G

R/G

109.0

81.80

B/G

R

B

G

R

G

120.0

90.00

110.0

R

B

G

 B/R 

R/G

75.00

91.70

 G/R

 B/G

Analog display (R/G, B/G)

Analog display (

R)

Analog display (B/R, G/R)

Содержание CA-310

Страница 1: ...Display Color Analyzer CA 310 INSTRUCTION MANUAL Before using this instrument please read this manual...

Страница 2: ...g and digital interfaces must be certified to the respective IEC standards i e IEC 60950 for data processing equipment Furthermore all configurations shall comply with the system standard IEC 61010 1...

Страница 3: ...ds Doing so may cause electric shock Do not disassemble or modify the CA Series Do ing so may cause a fire or electric shock Take special care not to allow liquid or metal ob jects to enter the CA Ser...

Страница 4: ...changes of temperature to avoid condensation The instrument must not be used in areas where there is an excessive amount of dust or where the humidity is excessively high The instrument should be use...

Страница 5: ...ntain measurement accuracy of instrument For details on checkup please contact the nearest Konica Minolta authorized service facility Disposal Method Please dispose of this product and the packing mat...

Страница 6: ...e optional 4 Probe Expansion Board CA B15 is used 27 3 Connecting the Power 28 4 Inputting the Vertical Synchronizing Signal 28 Turning the Power ON OFF O 29 1 Turning the Power ON OFF O 29 2 Instrume...

Страница 7: ...ent Section _____________________________________________75 Measurement 76 1 Performing Measurement 76 2 Holding the Measured Values 77 3 Displaying the Measured Values 78 White Balance Adjustment in...

Страница 8: ...e zero point adjustment method Measurement cannot be performed if zero calibration is not completed The Preparation Setting Measurement section explains the procedure up to measurement The Measurement...

Страница 9: ...tion Gives detailed explanation of user calibration and explains its execution method Target color is also set at this time Analyzer Mode Gives detailed explanation of ana lyzer mode and explains how...

Страница 10: ...ains how to adjust white balance Page 81 This section explains communication with PC via RS 232C or USB Page 88 Remote Measurement Explains how to perform measurement from the PC remotely Page 88 Comm...

Страница 11: ...Please read when an error message appears in the LCD display section Troubleshooting Guide Please read when the instrument does not function correctly Specifications Measurement Quick Guide Provides...

Страница 12: ...l input terminal 10 Probe connector P1 15 4 Probe Expansion Board slot 11 USB connector 12 RS 232C connector 14 AC power connector Rear 2 Digital display 1 POWER switch 4 Measurement mode indications...

Страница 13: ...Flicker Measuring 27 Probe or LED Flicker Measuring 10 Probe is connected E LCD display Displays the memory channel probe no ID name warning and settings F HOLD LED Lights up during hold G REMOTE LED...

Страница 14: ...ibration set this ring to the 0 CAL position to block entry of light into the probe For measurement Set the ring to the MEAS position to perform measurement C Ring stopper Stops the ring at two positi...

Страница 15: ...connectors on the 4 Probe Expansion Board before measurement Location of explanations Connecting method Page 26 Measuring method Measurement Preparation Setting Measurement sections P2 P3 P4 P5 Probe...

Страница 16: ...dependability measurement reproducibility will be higher if the installation angle of the measuring probe is kept constant for all measurements Use of the hood standard accessory not only prevents en...

Страница 17: ...AS 0 CAL To turn the ring the stopper must be pulled toward you to unlock it MEAS To perform measurement the ring must be set in this posi tion 0 CAL To perform zero calibration the ring must be set i...

Страница 18: ...r LED Universal Measuring 10 Probe CA PSU32 35 is connected Analyzer mode xyLv T uvLv RBG R G B G G RBG R B G G R u v Lv XYZ xyLv When LED Flicker Measuring 27 Probe CA P32 35 or LED Flicker Measuring...

Страница 19: ...9 2 ALPHA key Used to enter alphabets This key enables you to use the number key as alphabet keys Pressing this key again will restore the origi nal function of the ten key 3 Alphabet keys Used to ent...

Страница 20: ...leeping sound A whistling sound turned on next time is heard when the setting is saved 3 Holding down the MR key When xyLv T uvLv or XYZ measurement mode is selected for two to four seconds Displays s...

Страница 21: ...0 Probe CA PS32 35 is connected Analyzer mode xyLv T uvLv RBG R G B G G RBG R B G G R u v Lv Flicker XYZ xyLv 2 Digital display section Displays the measured values When xyLv measurement mode is selec...

Страница 22: ...ed For R reference R B R and G R are displayed When flicker mode is selected Flicker is dis played 4 LCD display section Displays the memory channel probe no ID name warning and set tings In case of e...

Страница 23: ...the measured value has Digital display exceeded 999 9 Analog display Not lit LCD display FLICKER ERROR OVER When Lv luminance is the following Digital display under 0 1cd m2 for LED Flicker Measuring...

Страница 24: ...22...

Страница 25: ...tal conditions for the instrument and notes on installation About Connection Explains how to connect measuring probes and connect the power cord It also explains installation method of the optional 4...

Страница 26: ...cted inside the instrument do not use the instrument Doing so may result in a fire For periodic inspection contact a Konica Minolta authorized service facility Always use theAC power cord supplied as...

Страница 27: ...n poorly ventilated areas Do not use the instrument in a place where the temperature changes rapidly since measured values will be incorrect The instrument must not be used in areas where there is an...

Страница 28: ...ll the probe by holding the plug Never pull the probe by its cord Notes when Connecting the Probe Never connect or remove the measuring probe while the POWER switch is ON When connecting disconnecting...

Страница 29: ...cted LED Universal Measuring 27 Probe CA PU32 CA PU35 LED Universal Measuring 10 Probe CA PSU32 CA PSU35 LED Flicker Measuring 27 Probe CA P32 CA P35 LED Flicker Measuring 10 Probe CA PS32 CA PS35 A t...

Страница 30: ...nstrument to allow synchronous measure ment when EXT SYNC mode is selected However if another SYNC mode is selected it is not necessary to input the vertical synchronizing signal Connect the BNC cable...

Страница 31: ...R switch to ON first then turn ON the power to the external equipment Turning the Power OFF If the instrument is connected to external equipment turn OFF the power to the external equipment first then...

Страница 32: ...rd data 13 Luminance unit Page 32 cd m2 Changing the Instrument Status at Power ON Change necessary parameters and press the key for more than five seconds A bleep will sound followed by a whistling s...

Страница 33: ...onfirm the set tings Changing Method for parameter 12 13 For the setting method refer to the page given in the above table About the REMOTE Key The REMOTE key should not be pressed unless you are goin...

Страница 34: ...will be added one after another as shown 2 Press the key before a total of sixteen asterisks appear Keeping the key held down will display as shown below switching the luminance unit from one to anot...

Страница 35: ...ins the zero point adjustment method Measurement cannot be performed if zero calibration has not been completed Selecting Masurement Speed SYNC Mode Measurement Mode and the Number of Display Digits E...

Страница 36: ...on to take more time When the optional 4 Probe Expansion Board CA B15 is used Zero calibration will be performed simultaneously with all the connected measuring probes Operating Procedure Before start...

Страница 37: ...Zero Calibration Check Method If you want to check whether zero calibration has been performed correctly block entry of light into the measuring probe s receptor using a blackout curtain etc If the me...

Страница 38: ...LED Universal Measuring 10 Probe FAST SLOW When Lv drops below 12 0cd m2 CA PSU32 35 SLOW FAST When Lv exceeds 18 0cd m2 LED Flicker Measuring 27 Probe FAST SLOW When Lv drops below 2 0cd m2 CA P32 35...

Страница 39: ...ame input RANGE Measurement Speed Number of Digits Calibra tion standard RS232C Baud Rate PROBE 3 Pressthe keytodisplaythedesiredmeasure mentspeed Each time the key is pressed the measurement speed sw...

Страница 40: ...key to open the SYNC selection screen Each time the key is pressed the screen will switch in the order PROBE SYNC ID Name input RANGE Measurement Speed Number of Digits Calibration standard RS232C ba...

Страница 41: ...CA PU32 35 CA P32 35 or 0 3 cd m2 or higher for Measuring 10 probe CA PSU32 35 CA PS32 35 If Lv is lower than mentioned above 3 5 measurements sec spplies 2 The measurement speed for EXT and INT when...

Страница 42: ...lay section shows x y and Lv Analyzer mode G reference Used to display meeasurement RBG emission intensities as a percentage of the RGB emission intensities target color W Analog display shows measure...

Страница 43: ...mode will be selected automatically when the POWER switch is set to ON x y Lv x y Lv T Lv uv x y Lv R B G R G B G G R B G R B R G R u Lv v x y Lv x y Lv xyLv mode T uvLv mode Analyzer mode G Standard...

Страница 44: ...pressed the number of dis play digits switches alternately between 4 FIGURES and 3 FIGURES 4 Press the key to confirm the selection By default factory setting the instrument is set so that 4 FIGURES w...

Страница 45: ...Calibration standard RS232C Baud Rate PROBE 3 Press the key to display the probe no you want to select Each time the key is pressed the probe no switches in the order P1 P3 P5 P1 The probe type is di...

Страница 46: ...44...

Страница 47: ...and explains its execu tion method Target color is also set at this time Page 51 Analyzer Mode Gives detailed explanation of ana lyzer mode and explains how to input the RGB emission charac teristic f...

Страница 48: ...or User calibration must be performed in the following cases However it is not possible to perform user calibration using the memory channel CH00 To correct variation of readings that occur due to the...

Страница 49: ...ds for setting changing the target color 1 Setting changing the target The display s measured values are set as the target color color by measurement This method can be used for any memory channels 2...

Страница 50: ...es the RGB emission characteristic for analyzer mode is stored in their common memory irrespective of measurement mode Therefore when matrix calibration is performed the RGB emission characteristic fo...

Страница 51: ...tting the RGB emission When you select analyzer measurement mode and input the RGB characteristic for analyzer mode emission characteristic for analyzer mode also set the target color page 59 W Since...

Страница 52: ...00K and 9300K 4 Press the key to confirm the selection The selected calibration standard will be set for CH00 as well as for all the memory channels that have not been user calibrated By default facto...

Страница 53: ...probes are used At the same time as user calibration is performed for a memory channel the obtained color will be set as the target color to that memory channel The target color is the color used as t...

Страница 54: ...alibration will be set Operating Procedure When the optional 4 Probe Expansion Board CA B15 is used Select the probe no to be white calibrated White calibration can be performed independently for each...

Страница 55: ...be set as the target color when the correction factor is entered 10 Press the HOLD key to start measurement To cancel white calibration press the key before pressing the key at step 9 To view the whit...

Страница 56: ...r mode is to be set the previous RGB emission characteristic will be deleted and the WRGB set for matrix calibration will be set as the RGB emission characteristic Operating Procedure When the optiona...

Страница 57: ...ite calibration see step 7 in Performing White Calibration on page 52 5 Press the key The LCD display section will return to the user calibration input screen with the mark displayed on the left of R...

Страница 58: ...8 will be set as the target color when the correction factor is entered Steps 5 to 8 can be performed in any order Pressing the or key before pressing the key at step 9 allows you to re enter the emis...

Страница 59: ...n mode d Matrix calibration with Konica Minolta s calibration standard 6500K h Matrix calibration with Konica Minolta s calibration standard 9300K a White calibration user calibration m Matrix calibra...

Страница 60: ...ion intensity of R B and G monochrome lights and the target color W are set and measurement is performed in analyzer mode Digital display section R B G Outputs of the currently measured monochrome lig...

Страница 61: ...teristic for analyzer mode Operating Procedure When the optional 4 Probe Expansion Board CA B15 is used Select the probe no for which the RGB emission characteristic for the display is to be input The...

Страница 62: ...ppear on the left of W 8 Press the key The RGB emission characteristic for the display s analyzer mode and target color W will be set Steps 4 to 7 can be performed in any order Pressing the or key bef...

Страница 63: ...e sure that the measuring probe is placed against the display If it is tilted or moved it is not possible to input accurate emission characteristic Take care not to let the measuring probe be exposed...

Страница 64: ...CH00 When you want to perform measurement using Konica Minolta s calibration standard without user calibration and want to use the analog display function When you want to set a color that differs fr...

Страница 65: ...default factory setting the instrument is set so that P1 will be selected automatically when the POWER switch is set to ON 1 Press the MEMORY CH and keys until the memory channel where you want to se...

Страница 66: ...may result in breakdown or wire break age If OVER is currently displayed it is not possible to set the currently measured color as the target color since the instrument s measurement range is exceeded...

Страница 67: ...ches in the order P1 4 Press the key to confirm the selection By default factory setting the instrument is set so that P1 will be selected automatically when the POWER switch is set to ON 1 Press the...

Страница 68: ...key to enter the Lv value 4 Press the key The target color will be set to CH00 To cancel target color setting press the key before pressing the key To view the target color you set press the MR key Fo...

Страница 69: ...e of display with what colors Operating Procedure Probe no CH01 EXT Ad P1A MENU SELECT PUSH SPACE KEY CH01 EXT Ad P1 _ Memory channel 1 Press the MEMORY CH and keys until the memory channel to which y...

Страница 70: ...position 5 Press the key The ID name will be set for the selected memory channel To cancel ID name setting press the key Notes when Setting an ID Name The ID name will be kept even if the POWER switc...

Страница 71: ...licker Mode Flicker value The range must be set independently of 1 2 and 3 In the case of 1 the range set here will be used commonly by all the modes Thus for instance if x and y are set to 2 and Lv i...

Страница 72: ...rence Rt R Rt 100 R 100 R B R 100 R B R G R R G R 100 where Gt Rt Target color values being 100 R G B Measurement Values For flicker mode Measured values are displayed as they are The display lights u...

Страница 73: ...s set to 2 0 1 Press the and then key to set the x y range 2 Press the key The cursor _ will move to the Lv position 3 Press the and then key to set the Lv range 5 Press the key The ranges will be set...

Страница 74: ...wing keys during range setting Doing so will cancel range setting and activate the mode corresponding to the pressed key 0 CAL MODE REMOTE MR MEMORY CH Digital and Analog Display In the case of four d...

Страница 75: ...ue that was entered when the previous user calibration was performed The values for the target color will be displayed if user calibra tion has not been performed When matrix calibration is performed...

Страница 76: ...to input the analyzer mode RGB emission characteristic or set the target color for the currently selected memory channel will be displayed When the optional 4 Probe Expansion Board CA B15 is used The...

Страница 77: ...section explains measuring methods From the Settings Section Measurement Explains measuring methods how to hold the measured values and how to read them White Balance Adjustment in Analyzer Mode Expla...

Страница 78: ...5 or LED Flicker Measuring 10 Probe CA PS32 35 is connected for a long period of time perform zero calibration approximately every hour Make sure that the measuring probe is placed straight against th...

Страница 79: ...measurement results All measurement values can be acquired by communicating with PC how ever the display of the main unit always displays measurement values 3 to 5 times second and does not display a...

Страница 80: ...color in the case of monochrome light G When analyzer mode R standard is selected R Difference from the target color in the case of mono chrome light R G R B R Ratio of measured values For details on...

Страница 81: ...Digital display section Display contents Contrast flicker value Display range 0 0 to 999 9 up to the first decimal place Analog display section Display contents Contrast flicker value For details on t...

Страница 82: ...ccurred due to shift of the zero point because the ambient temperature has changed since zero calibration Action Perform zero calibration Measurement can still be performed even if E2 is currently dis...

Страница 83: ...B and G in the digital display section when measurement is performed in analyzer mode this indicates that the color of the display measured is the same as the target color W i e the xyLv values are t...

Страница 84: ...ng to the MEAS position and start measurement 4 Press the MODE key to select analyzer measure ment mode RGB 5 Press the MEMORY CH and keys to select the memory channel for which the RGB emission chara...

Страница 85: ...value for B changes from 81 80 to 100 0 When all the values for R B and G are changed to 100 0 adjustment of the white color of the display to the target color W i e the xyLv values are the same is co...

Страница 86: ...84...

Страница 87: ...PC via RS 232C Communicating with PC via USB Explains how to connect the USB cable to enable communication with PC via USB Remote Measurement Explains how to perform measurement from the PC remotely I...

Страница 88: ...iring diagram Connecting Method 1 Set the POWER switch to OFF O 2 Connect the instrument to the computer with the RS 232C cable 3 Connect the cable to the connector and secure them with two screws fir...

Страница 89: ...instrument is set so that 38400 will be selected automatically when the POWER switch is set to ON To cancel selection of RS 232C baud rate press the key Notes when Selecting the RS 232C Baud Rate The...

Страница 90: ...USB port on the instrument If the computer has two or more USB ports the USB cable can be connected to any of them When used at the same time with other machines sometimes it will not work normally N...

Страница 91: ...ease read when an error message appears in the LCD display section Troubleshooting Guide Please read when the instrument does not function correctly Specifications Measurement Quick Guide Provides an...

Страница 92: ...mperature detection section are digitized in the A D conversion section A D conversion is per formed simultaneously within the measurement time according to SYNC mode D The digitized values counts are...

Страница 93: ...d to express colors that are slightly off the blackbody locus When a color is on the isotemperature line the color temperature at the point where that line crosses the blackbody locus is assumed to be...

Страница 94: ...one of the user calibration methods white calibration single color calibration or matrix calibration RGB W calibration These user calibration methods have the following features so the user calibratio...

Страница 95: ...ards the display s R G and B Thus even if R monochrome light is emitted on the display s screen an output will be provided from each sensor x y z This is also true in the case of G and B monochrome co...

Страница 96: ...s are emitted by the display at the same time R G and B are displayed in percentage to each monochrome light of the target color W being 100 Therefore before performing measurement in analyzer mode th...

Страница 97: ...each portion is received by a sensor that has a spectral sensitivity similar to the CIE 1931 color matching function Three sensors in total IEC 61747 6 stipulates the following measuring requirements...

Страница 98: ...e as that of the display s vertical synchronizing signal Flicker Measurement Method Two kinds of quantifying measurement methods are available contrast method and JEITA method With the CA 310 with LED...

Страница 99: ...n addition to the AC DC component ratio that is defined by the contrast method With the JEITA method the measuring devices shown below are required for measurement Fig 3 shows that the output signal F...

Страница 100: ...he maximum value among all the frequency components Px1 Px2 in the case of Fig 4 except for P0 that is the component of frequency 0 will be set as Px With the JEITA method the flicker amount in this e...

Страница 101: ...5 3 is the figure of emission characteristics of popular displays Fig 5 1 Emission characteristics of CRT Fig 5 2 Emission Characteristics of PDP Fig 5 3 Emission Characteristics of Active matrix LCD...

Страница 102: ...ot vary in a frame scanning period as shown in Fig 5 3 So inaccurate measurement data is sometimes obtained for CRT or PDP with the CA P32 35 or CA PS32 35 The emission intensity of some types of LCD...

Страница 103: ...o the following points may result in death or serious injury If you are not going to use the instrument for a long time disconnect the AC power cord from the AC outlet Dirt or water may accumulate on...

Страница 104: ...000 2000 or 5000 47 5 47 5 ISO screw 5mm depth 6mm ISO screw 5mm depth 6mm ISO screw 5mm depth 6mm Tripod screw depth 6mm When used with the hood When used with the hood ISO screw 5mm depth 6mm Tripod...

Страница 105: ...easurement mode RGB is selected 1 The RGB emission characteristic for analyzer mode has not been input for the selected memory channel since shipment from factory 2 The currently used measuring probe...

Страница 106: ...connector P1 properly Connect the probe to the probe connec tor P1 properly Before connecting disconnecting the measuring probe make sure that the POWER switch is set to OFF A measuring probe was con...

Страница 107: ...9 Aborts selection of the number of display digits 0 Aborts calibration standard selection 1 Aborts RS 232C baud rate selection The instrument operates as follows if the error message 5 is displayed 1...

Страница 108: ...connected mea suring probe Cause The RGB emission characteristic for the display s analyzer mode has not been input Action Set the RGB emission characteristic for the display s analyzer mode If you wa...

Страница 109: ...tion does not end ZERO CALI BRATION is displayed in the LCD display section TOO BRIGHT is displayed even if the light is blocked properly Is the cable for the vertical synchronizing signal connected t...

Страница 110: ...EMOTE LED lights up 86 Is the created program correct Check it by referring to a sample program 10 During com munication with USB Data output from the instru ment cannot be imported to PC Commands or...

Страница 111: ...Install it correctly 27 Are the measuring probes connected to the probe connectors P2 to P5 properly Connect necessary number of probes to the probe connectors properly 26 27 14 Probe cannot be chang...

Страница 112: ...synchronization frequency 40 to 200 Hz Memory channel 100 channels Analyzer function Standard function Interface USB RS 232C 38 400 bps or below Multi point Measurement Max 5 points Use 4 Probe Expan...

Страница 113: ...urement Vertical synchronization frequency 40 to 200 Hz Memory channel 100 channels Analyzer function Standard function Interface USB RS 232C 38 400 bps or below Multi point Measurement Max 5 points U...

Страница 114: ...Vertical synchronization frequency 40 to 200 Hz Luminance or chromaticity measurement 40 to 130 Hz Flicker measurement Memory channel 100 channels Analyzer function Standard function Interface USB RS...

Страница 115: ...surement Vertical synchronization frequency 40 to 200 Hz Luminance or chromaticity measurement 40 to 130 Hz Flicker measurement Memory channel 100 channels Analyzer function Standard function Interfac...

Страница 116: ...until the desired SYNC mode appears 4 Press the key to confirm the selection To use EXT mode the vertical synchronizing signal used for the display must be input to the instrument Page 28 EXT UNIV IN...

Страница 117: ...probe No will be switched to the selected one 4 Press the key to confirm the selection Menu selection screen Press the key until the desired probe no appears PROBE selection screen When the optional 4...

Страница 118: ...ker measurement 1 Setting Changing the Target Color by Measurement Page 63 This method can be used for any memory channels HOLD LED Memory channel Probe no When the optional 4 Probe Expansion Board CA...

Страница 119: ...ng Section Pages 45 to 74 Cursor To change the target color after user calibration 1 Setting Changing the Target Color Page 116 To set an ID name 2 Setting an ID Name Page 118 To use the analog displa...

Страница 120: ...e display which is now emitting monochrome light of B 2 Press the key 7 Input the emission characteristic of W 1 Place the measuring probe against the display which is now emitting monochrome light of...

Страница 121: ...RESULTING FROM THE MISUSE MISHANDLING UNAUTHORIZED MODIFICATION ETC OF THIS PRODUCT OR FORANY INDIRECT OR INCIDENTAL DAMAGES INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS INTERRUPTION OF BUSIN...

Страница 122: ...BGLFPK Printed in Japan 9222 A3CT 11 2010 2016 KONICA MINOLTA INC...

Отзывы: