60
1
. Press the MODE key to select analyzer measurement
mode (RGB).
2
. Press the MEMORY CH
and
keys until the
memory channel where you want to input the RGB emis-
sion characteristic appears.
3
. Press the
key.
The LCD display section will switch to the analyzer mode RGB emis-
sion characteristic input screen.
4
. Input the emission characteristic of R.
1
Press the measuring probe against the display, which is now emit-
ting monochrome light of R.
2
Press the
key. In the LCD display section, the “*” mark will
appear on the left of “R”.
5
. Input the emission characteristic of G.
1
Place the measuring probe against the display, which is now emit-
ting monochrome light of G.
2
Press the
key. In the LCD display section, the “*” mark will appear on the left of “G”.
6
. Input the emission characteristic of B.
1
Place the measuring probe against the display, which is now emit-
ting monochrome light of B.
2
Press the
key. In the LCD display section, the “*” mark will
appear on the left of “B”.
7
. Enter the target color (W)
1
Place the measuring probe against the display, which is now emit-
ting the target color(W).
2
Press the
key. In the LCD display section, the “*” mark will
appear on the left of “W”.
8
. Press the
key.
The RGB emission characteristic for the display’s analyzer mode and target color (W) will be set.
* Steps 4 to 7 can be performed in any order.
* Pressing the
,
,
or
key before pressing the
key allows you to re-enter the emission characteristic.
* To cancel emission characteristic setting, press the
key before pressing the
key.
* To change the target color you set, change it as explained in “1. Setting/Changing the Target Color by Measurement” (page 63).
Even if the target color is changed, the currently set RGB emission characteristic for display’s analyzer mode will remain unchanged.
* Pressing the MR key displays “100.0” as the target color value for R, B and G.
<Error Messages in LCD Display Section>
… For other error messages, refer to page 103.
●
“E1”
•
Cause
1
: The display’s RGB emission characteristic has never been input for the currently selected
memory channel since shipment from the factory.
Action
: This error will disappear if you enter the emission characteristic.
•
Cause
2
: The currently used measuring probe is different from the
one that was used to input the display’s RGB emission
characteristic and target color (W).
Action
: Connect the same probe as the one used to input the
display’s RGB emission characteristic and target color (W).
Memory channel
CH01 EXT Ad P1A
E1 [ ]
CH01 EXT Ad P1A
[ ]
CH01 ALZ
P1 W R G B