KONGSBERG K-Sync
Condition
Resolution
The plot for a line is steady high
(transmitting) instead of showing the
states
The ping rate for the system is higher than
the trigger display resolution (50 ms),
display will not show individual pulses
correctly unless ping rate is 10 Hz or
below.
There is no red pulse indicating transmit
in the trigger display for some of the echo
sounders
Trigger pulse will only be shown if the
echo sounder provides the
transmitting
signal
and the signal is enabled in the
installation settings.
Echo sounder state toggles between
standby and active.
Check that the installation parameters for
ready to transmit are correct (correct input
terminal? Correct signal active level?)
Echo sounder goes to standby mode
inadvertently and stops pinging when
surveying in deep water.
Check that the ready to transmit timeout
parameter in the installation settings for
the echo sounder is long enough for the
current survey depth.
Display data degraded LED frequently
blinking.
• High CPU usage? (> 40%).
• If network connection is shared, is
there significant competing network
utilization? (usage > 10%).
54
342435/B