EM 122
Note
Dismount the Mounting Frame and transport it to a locksmith shop and use a
milling machine.
3
Transfer the hole-pattern according to the drawing. Drill and cut the thread.
4
Mounting frame to be checked with all baffle plates before re-installing.
5
Re-install the mounting frame according to the instructions in the procedure
on page 30.
How to mount the RX frames
The mounting frames must be handled with care. Please observe normal safety
precautions for dockyard work and welding.
1
Mount the baffle plates to the mounting frames. Use moderate torque. And secure
with Loctite 242
Note
The baffle plates must be installed with correct orientation!
• The number of baffle plates depends on the length of the array
• For easier handling is a text and a arrow mark on every baffle plate.
2
The mounting frame must be bolted onto the flat bars inside the casing.
Note
The mounting frame must be installed with correct orientation!
• The number of flat bars depends on the length of the array
• There are two (2) holes in each flat bar
3
The frame must be bolted properly to the bars with torque of approximately 187 Nm.
4
Check that the frame is mounted completely flat.
• No point on the frame may deviate from the optimal plane with more than
+-1.6mm
5
Repeat step 4 until the alignment procedure has been performed successfully.
• Refer to
on page 96 chapter for a proposed measuring method.
Refer to
on page 109 for mounting frame principles.
How to mount the TX frames
1
The mounting frame must be bolted onto the flat bars inside the casing.
• The number of flat bars depends on the length of the array.
• There are three (3) holes in each flat bar.
– The 0.5 degree mounting frame is a two-part type, that has to be assembled
before bolting onto the flat bars inside the casing. Refer to drawing file on
page xxx for 0.5 deg mounting frame
30
317669/B
Содержание EM 122
Страница 1: ...EM 122 Multibeam echo sounder Installation Manual...
Страница 2: ......
Страница 13: ...EM 122 Figure 1 System diagram 317669 B 11...
Страница 26: ...EM 122 Figure 6 EM 122 Gondola example Cd021 107 24 317669 B...
Страница 27: ...Transducer arrays Figure 7 EM 122 blister and steel pipe arrangements example 317669 B 25...
Страница 42: ...EM 122 B B B Plastic TX protection window 40 317669 B...
Страница 46: ...EM 122 Figure 10 EM 122 Transceiver unit mounting 44 317669 B...
Страница 52: ...EM 122 Figure 13 Rear view 50 317669 B...
Страница 54: ...EM 122 Figure 15 Preamplifier Unit mounting 687 52 317669 B...
Страница 60: ...EM 122 Figure 18 Remote Control Junction Box outline dimensions 58 317669 B...
Страница 85: ...Cable layout and interconnections Figure 32 Transceiver Unit sub racks rear view 317669 B 83...
Страница 88: ...EM 122 Figure 34 Preamplifier underneath view 86 317669 B...
Страница 97: ...Cable layout and interconnections Figure 40 EM 122 Transceiver Unit s rear panel RX 317669 B 95...
Страница 126: ...EM 122 Figure 57 Cable layout 0 5 degree TX page 1 of 2 Cd021 106 Page 1 of 2 315396 Pro03 124 317669 B...
Страница 127: ...Drawing file Figure 58 Cable layout 0 5 degree TX page 2 of 2 Cd021 106 Page 2 of 2 315396 Pro03 317669 B 125...
Страница 128: ...EM 122 Ice window Figure 59 TX Ice window 126 317669 B...
Страница 129: ...Drawing file Figure 60 RX Ice window 317669 B 127...
Страница 131: ...Drawing file Preamplifier Unit Figure 62 RX Preamplifier Unit internal cabling 317669 B 129...
Страница 132: ...EM 122 Figure 63 RX Preamplifier Unit internal cabling 130 317669 B...
Страница 154: ...2013 Kongsberg Maritime...