qb701 User
’s
Manual
21
Auto shooting mode can be activated in both Clock mode and
Multipoint mode.
When device is in Multipoint mode and enough readings are
collected
–
press
to proceed to the result screen.
When device is in Clock mode and three/four readings are taken
device will automatically proceed to the result screen.
READINGS QUALITY ASSESSMENT
While taking readings device performs assessment of
data quality based on standard deviation. Quality is
indicated by color of dots at measurement positions:
Blue
–
assessment is impossible (too few points
collected)
Green
–
good quality
Yellow
–
acceptable quality
Red
–
poor quality, must be re-measured.
The factors which may affect readings
–
high vibration, partial clipping of the laser beam,
mechanical looseness, accidental alteration of transducers position (e.g. by touching it).
Readings quality assessment is useful option which helps to detect such conditions.
EDITING OF MEASUREMENT POINTS
It is possible to edit collected data when poor quality
data is detected.
Good
Poor
Acceptable
Poor