K41091 Micro Carbon Residue and Ash Tester 

Operation and Instruction Manual 

 

K41091-Manual

    

 

-

7-

 

6. 

Basic Functions of the Apparatus: 

 
The apparatus has been designed to perform tests 
for both carbon residue and ash. 
 

WARNING:

  It  has  been  determined  that  at  the 

higher  temperatures  of  775°C  for  the  Ashing  test, 
the  temperature  stability  of  the  furnace  is  outside 
the acceptable range of testing standards.  
 
Two  preprogrammed  test  procedures  have  been 
developed and implemented into the instrument.  
 
 

Program No. 1: Micro Carbon Residue 

 

Start T (°C) 

End T

 

(°C)

 

Duration 

(min) 

Gas Flow 

50 

50 

10 

NIT 

50 

500 

36 

nit 

500 

500 

15 

nit 

500 

50 

NIT 

 

Program No. 2: Ash 

 

Start T (°C) 

End T

 

(°C)

 

Duration 

(min) 

Gas Flow 

50 

50 

10 

NIT 

50 

775 

60 

NIT 

775 

775 

120 

AIR 

775 

50 

AIR 

 

Where: 
 

Air = Low Flow Air 
nit = Low Flow Nitrogen 
AIR = High Flow Nitrogen 
NIT = High Flow Nitrogen 

 
During a test procedure, the instrument follows the 
temperature  profile  of  the  program  selected  and 
also  manages  the  gas  flow  (air  or  nitrogen) 
following the sequences defined in the program.  
 

NOTE:

  The  gas  flow  rate 

(High  and  Low  Flow)

 

must be adjusted and set manually before starting 
the test.  
 
Flow meter Calibration Scale as per ASTM D4530 
 

30mm = 150mL/min 

120mm = 600mL/min 

 
NOTE:

  It  is  possible  to  follow  the  temperature  of 

the furnace by connecting an External PC to Serial 
Line  1  and  using  the  Data  Acquisition  Software 
provided with the accessories of the instrument.  
 

NOTE:

 The calculation  of residue percentage  can 

also be determined by the apparatus. Inputting the 
initial  and  final  weight  of  the  samples  into  the 
software will then calculate the residue percentage 
automatically.  
 

5.2  Starting a Test 

 

WARNING:

 

For  either  program  selected,  the 

Instrument  must  be  placed  under  a  hood  to 
evacuate  the  fumes  generated  by  the  furnace 
during testing. When performing an Ashing Test, it 
is  highly  recommended  to  reduce  the  light 
distillation  products  in  the  sample  by  using 
nitrogen  to  avoid  any  risk  of  spontaneous 
combustion. 

Rapid  Combustion  must  not  be 

performed in the Furnace!

 

 

WARNING:

  It  is  essential  to  understand  the 

standard test method of the test to be performed.  
 

WARNING:

 It will be necessary to work with a hot 

furnace  and  samples  before  and  after  each  test 
run. Therefore, it is required for the operator to use 
Personal  Protective  Equipment  (PPE)  and  the 
suitable accessories provided with the apparatus.  
 

WARNING: 

When conducting an Ashing Test it is 

required to use quartz vials (sold separately). The 
vials  provided  with  the  instrument  are  of 
borosilicate  glass  and  are  intended  for  only  the 
Micro Carbon Residue Test.  
 

1. 

Check that the condensate tank is empty.  

 
WARNING:

 The oven temperature must be below 

100°C before checking the condensate tank.  
 

 

 

Figure 5: Condensate Tank Check 

 

2. 

Remove the Insulating Lid 

 
 
 
 
 

Содержание K41091

Страница 1: ... Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K41091 Micro Conradson Carbon Residue and Ash Tester ...

Страница 2: ......

Страница 3: ...scriptions Back 5 5 Operation 6 5 1 Instrument Start up 6 5 2 Starting a Test 7 5 3 Control Panel Operation 8 5 3 1 Creation of New Programs 8 5 3 2 Software Configuration Functions 9 5 3 3 Hardware Configuration Functions 10 6 Maintenance 11 6 1 Maintenance Checks 11 6 2 Routine Maintenance 11 6 2 1 Cleaning of Furnace Opening and Hot Lid 11 6 2 2 De Carbonizing the Furnace 11 6 2 3 Temperature C...

Страница 4: ...ehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical support Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 Email info koehlerinstrument com http www koehlerinstrument com 1 2 Recommended Resources and Pub...

Страница 5: ...ions The use of this equipment may involve hazardous materials and operations This manual does not purport to address all of the safety problems associated with the use of this equipment It is the responsibility of any user of this equipment to investigate research and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use Equipment...

Страница 6: ...artz sample vial re usable Ash Content Test K41000 5 16 ml Quartz sample vial re usable Ash Content Test K41000 8 4 ml Borosilicate glass sample tube MRC test 3 2 Unpacking Carefully unpack and place the instrument and accessories in a secure location Ensure that all items listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all packing...

Страница 7: ...rns off the power to the entire unit 4 Numeric Keypad For entering sample data and parameter values for testing 5 Key Function Buttons For navigation through the instruments sub menus Note The function of these buttons and subsequent display icons will change depending on the menu the user is in 6 Display For user friendly control of the instrument 7 Insulating Cover For enhanced temperature stabi...

Страница 8: ... an attempt to further evacuate the exhaust fumes The chimney must remain free to ensure optimal exhaust of the system Any connection on the chimney or its extension pipe would be likely to cause Condensates in the piping of the apparatus A backpressure which would disturb the test by modifying the flows and cause leaks on the hot lid of the furnace 2 Connect the line cord to properly fused and gr...

Страница 9: ...cting an External PC to Serial Line 1 and using the Data Acquisition Software provided with the accessories of the instrument NOTE The calculation of residue percentage can also be determined by the apparatus Inputting the initial and final weight of the samples into the software will then calculate the residue percentage automatically 5 2 Starting a Test WARNING For either program selected the In...

Страница 10: ...ntage ratio of residue remaining at the end of the test 13 Choose the desired program Micro Carbon Residue or Ash 14 Confirm the launching of the test by pressing Go once again on the display panel 15 The instrument will then sound a series of beeps The display will now indicate the temperature of the furnace the duration of the test the current segment of the test program running and the limiting...

Страница 11: ...e two 2 in the single digit place resulting in the desired value 7 Set Point Function This function allows the user to input the idle set point temperature of the furnace between tests This value is limited to the range from 20 to 80 C 8 Standby Function This function allows the user to shut down the apparatus via the control panel without having to switch off the circuit breaker or push the emerg...

Страница 12: ...rature of the furnace Either in degrees Celsius or Fahrenheit 3 Calibration Function This function allows the user to calibrate the furnace temperature probe 3 1 Physical Calibration This calibration is done by entering three 3 temperature points via the keyboard Enter these points by first replacing the furnace thermocouple with a temperature simulator To do this open remove the furnace thermocou...

Страница 13: ...e each test it is recommended to check the furnace exhaust The exhaust must be free from calamine If the apparatus has calamine deposits in the furnace exhaust it is recommended to remove the deposits using the cleaning cable provided 6 2 Routine Maintenance Depending on the frequency of use and the type of samples being tested it is recommended to perform routine maintenance procedures For exampl...

Страница 14: ...its Check the state of the chimney and chimney extension pipe for calamine deposits and clean in the same manner 6 2 3 Temperature Check of the Furnace To check the temperature of the furnace it is necessary for a control thermocouple to be in place In order to make this possible the hot lid has a small thermocouple port located along the outside If the actual temperature of the furnace does not c...

Страница 15: ...tension Pipe Gripping Hook Complete Furnace Furnace Thermocouple Safety Thermocouple Thermal Printer Paper for Thermal Printer Insulating Lid Small Vial Borosilicate Large Vial Borosilicate Long Vial Borosilicate Borosilicate Vial 12mm dia 72mm height Support for Porcelain Crucible Control Thermocouple w cable Furnace Thermocouple Calibration Cord Upstream Pressure Gauge Condensate Tank Joint for ...

Страница 16: ...K41091 Micro Carbon Residue and Ash Tester Operation and Instruction Manual K41091 Manual 14 7 Diagrams 7 1 Flow Meter Calibration Chart ...

Страница 17: ...K41091 Micro Carbon Residue and Ash Tester Operation and Instruction Manual K41091 Manual 15 7 2 Function Flow Chart ...

Страница 18: ...the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replacement of the product In no event shall the cost of the purchaser s remedy exceed the purchase price nor shall Koehler Instrument Company be liable for any special indirect incidental conse...

Страница 19: ...K41091 Micro Carbon Residue and Ash Tester Operation and Instruction Manual K41091 Manual 17 Notes ...

Страница 20: ...K41091 Micro Carbon Residue and Ash Tester Operation and Instruction Manual K41091 Manual 18 Notes ...

Отзывы: