K41091 Micro Carbon Residue and Ash Tester
Operation and Instruction Manual
K41091-Manual
-
8-
3.
Remove the Hot Lid with the provided hook
accessory.
Figure 6: Hot Lid Handling
4.
Check that the furnace exit is not clogged by
calamine. If this is the case please see the
maintenance section of this manual.
5.
Place the sample vial holder with desired
samples in the furnace with the hook
accessory provided.
NOTE:
Be cautious of the walls of the furnace as
the surface is maintained at an above ambient
temperature prior to testing.
Figure 7: Sample Vial Holder Handling
6.
Replace the Hot Lid with the hook accessory
to ensure a tight seal.
7.
Replace the Insulating Lid on the Furnace.
8.
Make sure the Gas Feeder Valves from the
Gas Source are Open.
9.
Check that the ventilation hood is running and
that the chimney extension pipe is properly
installed on the Furnace Chimney.
10.
Begin the test by pressing the “Go” key on the
display panel.
11.
Enter the ID number for each sample.
12.
If desired, enter the weights of the samples.
This allows for an automatic calculation of the
percentage (ratio) of residue remaining at the
end of the test.
13.
Choose the desired program (Micro Carbon
Residue or Ash).
14.
Confirm the launching of the test by pressing
“Go” once again on the display panel.
15.
The instrument will then sound a series of
beeps. The display will now indicate the
temperature of the furnace, the duration of the
test, the current segment of the test program
running, and the limiting values of the
temperature of the current segment.
Figure 8: Display Parameters during Test
16.
At the end of the test, the instrument will
indicate the status to the operator on the
display. Wait for this message before
removing the samples from the furnace.
NOTE:
Although the instrument authorizes the
user to remove the samples, it is necessary to be
extremely careful because of the high temperature
of the furnace and samples.
5.3 Control Panel Operation
5.3.1 Creation of New Programs
In addition to the Two (2) factory preinstalled
programs, the user may create custom programs.
Each program is comprised of 4 segments just as
the preprogrammed procedures are structured.
Each temperature segment is defined by its initial
temperature and its final temperature.
The user must then also define the time duration
of each segment as well as the type of gas and its
flow rate.