K2269X Low Temperature Pressure Viscometer 

Operation and Instruction Manual 

 

K2269X-Manual

    

 

-

8-

 

1.  Process  Temperature  Display.

  The 

upper red LED display shows the process 
temperature as read from the RTD probe. 

 

2.  Set  Point  Temperature  Display.

  The 

lower  green  LED  display  shows  the  set 
point temperature of the controller. 
 

3.  Programming  Key.

  Permits  scrolling 

through  controller  menu  parameters.  One 
Level Forward 
 

4.  Down  Key.

  Used  to  decrease  the  set 

point  temperature  and  to  decrease  or 
change 

parameter 

values 

when 

programming the temperature controller.  
 

5.  Up  Key.

  Used  to  increase  the  set  point 

temperature  and  to  increase  or  change 
parameter  values  when  programming  the 
temperature controller. 
 

6.  Exit  /  Function  Key.

  This  key  is  used  to 

exit or leave a level. One level backward  
 

IMPORTANT 

NOTE:

 

The 

digital 

temperature  controller  for  the  unit  comes 
pre-programmed from the Koehler factory. 
Please do NOT attempt to re-program the 
digital  temperature  controller  as  this  will 
void the product warranty. If assistance is 
required, please do not hesitate to contact 
the Koehler technical service department. 

 

Setting  the  Temperature

.  Set  the  desired 

operating  temperature  by  adjusting  the  set 
point with the up and down keys. The set point 
will  be  displayed  in  the  lower  green  Set  Point 
LED display and the actual temperature will be 
displayed  in  the  upper  red  Process  LED 
display.  Please  allow  the  instrument  to  fully 
equilibrate before proceeding with any testing. 

 

Temperature Calibration.

 This routine allows the 

digital  temperature  controller  to  be  calibrated 
to a certified thermometer.  

 

a.  Use  a  certified  calibrated  measuring 

device  to  acquire  the  temperature. 
Calculate  the  difference  between  the 
measuring  device  and  the  Process  value 
displayed on the controller. 

b.  Press the program key two times until 

PCt

 

is  displayed  in  the  lower  green  LED 
display.  Press  the  DOWN  key.  CAL  will 
display on the lower green display. If there 

is  a  value observed  in  the upper red LED 
display, add it to the calculated difference 
obtained  in  the  previous  step.  This  is  the 
offset value. 

 

c.  Press  the  Program  Key.  The  lower  green 

display will flash. Use the up or down keys 
to  adjust  to  the  new  calibration  offset 
value  on  the  upper  red  display  calculated 
in  the  previous  step.  When  the  value  has 
been 

entered, 

the 

controller 

will 

automatically  store  the  value.  The  lower 
green  display  will  stop  flashing.  If  further 
adjustments  are  necessary,  press  the 
Program  Key  again.  Resume  regular 
operations by pressing the Exit / Function 
key two times. Verify if the new calibration 
is  correct  by  observing  the  upper  red 
display  and  comparing  the  value  with  the 
calibrated reference device.  

 

Auto  Tune.

  This  routine  allows  the  digital 

temperature  control  to  learn  the  heating 
parameters needed for any particular set point 
temperature.  This  operation  should  be  done 
when  installing  a  new  unit,  after  replacing  or 
changing  the  bath  medium  type,  or  utilizing  a 
different  temperature  set  point  20%  different 
from 

the 

previously 

used 

set 

point 

temperature. 

 

a.  Set  the  operating  temperature  to  the 

desired setting. 

 
b.  Press  the  up  and  down  arrow  buttons 

simultaneously for about 5 seconds. When 
Auto  Tune  is  active,  the  lower  green  LED 
display  will  blink 

TUNE

.  Auto  Tune  will 

automatically toggle off when the set point 
temperature is reached. Auto tune can be 
terminated  by  pressing  the  up  &  down 
buttons simultaneously again.  
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание K2269X

Страница 1: ...enue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...tured in conformance with the applicable standards set forth in this certification Specifications ASTM D1092 This unit is tested before it leaves the factory to ensure total functionality and complian...

Страница 4: ......

Страница 5: ...tion 5 4 1 Instrument Descriptions 5 4 2 Top Assembly Description 5 4 3 Control Box Descriptions 6 4 4 Gear Box and Motor Descriptions 7 4 5 Temperature Controller Descriptions 7 5 Operation 9 5 1 Ref...

Страница 6: ...pport of your laboratory testing needs Our products are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument servi...

Страница 7: ...t may exhibit potential hazards Any user must be familiarized with the possible dangers before use We also recommend consulting the Material Data and Safety Sheet MSDS on each chemical reagent for add...

Страница 8: ...e unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your claim if requested When submitting...

Страница 9: ...ar pump gears drive motor required to power the hydraulic system Refer to section 4 4 for detailed descriptions on this section of the apparatus 4 2 Top Assembly Description Figure 3 Instrument Descri...

Страница 10: ...nect the grease cylinder to the hydraulic system This port is also used as the main connection point for the hydraulic system calibration by utilizing the Calibration Pipe Assembly See Figure 1 Item G...

Страница 11: ...n See Figure 4 item 13 Variable Speed Motor is coupled to the speed reducer and is capable of operating at two speeds required to simulate flow rates attained by 40 tooth gear and 64 tooth gear as per...

Страница 12: ...a certified calibrated measuring device to acquire the temperature Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two t...

Страница 13: ...ion instructions for the temperature controller NOTE Allow for the test chamber to cool and come to temperature equilibrium prior to turning on Drive Motor Recirculating Pump See Figure 4 Item 13 It m...

Страница 14: ...imed 11 Once a steady flow or constant drip rate is observed the hydraulic system is primed and free of air bubbles 12 Using the relief valve at the right hand side of the calibration piping assembly...

Страница 15: ...esting purposes If not attached ferrule will be in thermocouple packaging 3 Remove bottom cap from Grease Cylinder F 4 Invert grease cylinder and place Grease Cylinder Piston C inside Grease Cylinder...

Страница 16: ...cylinder loading procedure is now complete Before attaching a capillary changing the flow of the hydraulic fluid and starting a test the sample grease must attain temperature equilibrium with the test...

Страница 17: ...may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenan...

Страница 18: ...104 002 Fuse 25A Slo Blo 5X20 M M 220 240 013 Cartridge Heater 500W 240V 265 500 001 RTD Temperature Probe x 12 275 103 044 Temperature Controller 100 240V 288 230 018 Motor 208 230V 3 Phase 288 230 0...

Страница 19: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 15 7 Wiring Diagrams 7 1 220 240V 50Hz Unit Wiring K22695...

Страница 20: ...e the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the re...

Страница 21: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 17 Notes...

Страница 22: ...K2269X Low Temperature Pressure Viscometer Operation and Instruction Manual K2269X Manual 18 Notes...

Отзывы: