NSC
DT0307
Page
7
3. Turn slowly the “SENSITIVITY ADJUSTMENT screw” counter clockwise until
RED LED switches off.
4. Turn 360º more counter clockwise to avoid a critical adjustment. In case of
sticky products it is recommended to increase this adjustment.
The instrument is now adjusted. When product will reach the probe relay will be
activated and RED LED will light.
If we want the relay to work the REVERSE, put DIP-SWITCH 1 to “MIN”.
Check that no product remains on the probe when unloading the tank.