To avoid mistakes, asks again if you are sure that you
want to calibrate the unit. Press (OK) another time to
enter.
Pressing (+) we enter in High level calibration.
If when you adjust the high level, the liquid in the tank is
100 % of the capacitance, we will indicate it in the
display using the (+) and (-) keys. Confirm with (OK).
The microprocessor makes the operations to calibrate
the maximum level of the unit. This operation takes
some seconds.
The display shows this message to indicate NMC has
finished the calibration internal process.
When the liquid level will be in the minimum point, it will
indicate 000.0 % and when the liquid level will be in the
maximum, 100.0 %.
Automatically, the display shows the main menu.
Pressing ESC the unit return to read state.
Pressing (+) we can enter in output adjustment.
menu and simulation mode.
9.2 Output Adjustment
Pressing any key, you enter in the main menu.
Press (+) to enter in Output.