DT0305
Pressing (-), we enter in simulate output menu.
In this screen, using the (+) and (-) keys, the loop
current (2 wire) will change in steps of 1mA. With this
option, you can do current testes without generator.
Pressing (OK) return to menu show in point 2.
Pressing (ESC) return to read mode.
9.4 Language
Pressing any key, you enter in the main menu.
Press (+) to enter in Output.
Pressing (+) or (-) we can choose the language for the
display.
Language available: Spanish, English, German, Italian,
French, Portuguese, Catalan.
Pressing (OK) return to main menu in language
selected.