Kobold HND-P105 Скачать руководство пользователя страница 4

HND-P105 

page 4 

 

HND-P105 K05/0117 

5. Operating 

Principle 

The KOBOLD manual pressure measuring devices HND-P105 are highly precise, 
compact pressure measuring devices that can be used universally. In conjunction 
with the appropriate external pressure sensors, precise measurement results 
over the entire measuring range can be achieved. 
Various pressure sensors are available for a multitude of measuring tasks and 
special applications. The respective measurement task determines which 
combination is selected. Naturally, these first-rate KOBOLD-measuring units can 
display more than just pressure. All devices in this series allow for 
minimum/maximum value memory, hold function, automatic self-shut-off, or zero 
point offset entry for all connected pressure sensors.  
 
 
 

6. Electrical 

Connection 

6.1  Mains Operation With Power Supply 

Warning: When using a power supply please note that operating 

voltage has to be 10.5 to 12 V

DC

Do not apply overvoltage!! Cheap 12 V-power supplies often have 
excessive no-load voltage.   
We, therefore, recommend using regulated voltage power supplies. 
Trouble-free operation is guaranteed by our power supply HND-

Z002. 
Prior to connecting the power supply to the mains make sure that 
the operating voltage stated at the power supply is identical to the 
mains voltage. 

6.1.1 Connecting/Changing 

Sensors 

Do not use unsuitable sensors. Connecting other devices/sensors as specificated 
may cause a damage to the instrument and device/sensor! Switch off device 
before changing the sensor. 
Connect sensor before switching on the device, otherwise the sensor may not be 
detected correctly. When connecting the sensor the connector may not lock 
correctly. In such case take the plug not at the casing but at the buckling 
protection at the end of the plug. If plug is entered correctly, it will slide in 
smoothly. To disconnect sensor do not pull at the cable but at the plug (to open 
locking mechanism). 
 
 

Содержание HND-P105

Страница 1: ...Operating Instructions for Handheld Pressure Measuring Devices with External Pressure Sensors Model HND P105...

Страница 2: ...uration 7 7 6 Measuring Of Water Level Display Unit m 8 7 7 Calibration Services 9 7 8 Pressure Connection To The Sensors 9 7 9 Error And System Messages 10 7 10 The Serial Interface 11 8 Maintenance...

Страница 3: ...before shipping and sent out in perfect condition Should damage to a device be visible we recommend a thorough inspection of the delivery packaging In case of damage please inform your parcel service...

Страница 4: ...upply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load voltage We therefore recommend using regulated voltage...

Страница 5: ...make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be...

Страница 6: ...uring from then on will be displayed relatively to the set tare values When the tare function is activated the arrow Tara appears in the display To deactivate tare function press Tara for 2 seconds Pl...

Страница 7: ...on 7 5 1 Unit Choice Of The Display Unit Choose the desired display unit the referring unit is displayed by means of a functional arrow in the display The selection is permanently stored in the sensor...

Страница 8: ...r deviations Input is in the display unit 7 5 6 SCAL Adjusting Sensor Scale The scale of the measuring can be influenced by this setting factor is in displayed value measured value 1 Scal 100 Standard...

Страница 9: ...l dia of 4 mm to pressure port B Port A will not be used Pressure sensors HND PS01 PS02 and P03 allow for measurements of under pressure up to the entire over pressure measuring range by re plugging t...

Страница 10: ...thout battery wrong voltage or polarity Check power supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufactu...

Страница 11: ...nge unit 7 Read max value 204 Read display range decimal point 12 Read ID number 208 Read of channels 176 Read min measuring range 214 Read Scale adjustment 177 Read max measuring range 216 Read Offse...

Страница 12: ...ower off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7mm high for measuring values and for min max memories hold functio...

Страница 13: ...0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 350 0 mbar rel 0 1 mbar max 1 4 bar 0 70 C G male stainless steel HND PS09 0 1000 mbar abs 1 mbar max 4 bar abs HND PS10 0 3500 mbar abs max 14 bar abs...

Страница 14: ...face converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with lo...

Страница 15: ...ectrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic product...

Отзывы: