Kobold HND-P105 Скачать руководство пользователя страница 10

HND-P105 

page 10 

 

HND-P105 K05/0117 

7.9  Error And System Messages 

Display 

Meaning 

What to do? 

 

Low battery power, device will only continue 
operation for a short period of time 

Replace battery 

 

Battery empty  

Replace battery 

Mains operation without battery: wrong voltage 

Check power supply, replace it when 
necessary 

 

 

or 

Err.9

 

No sensor connected 

Switch off device and connect sensor 

Connected sensor or device defective 

If second sensor available, check if device 
is ok. Return defective device/sensor to 
manufacturer for repair 

Measured value far out of allowable range 

Check: pressure not within sensor range?  

No display or 

confused 

characters, 

device does 
not react on 

keypress 

Battery empty 

Replace battery 

Mains operation without battery: wrong voltage 
or polarity 

Check power supply, replace it when 
necessary 

System error 

Disconnect battery and power supplies, 
wait shortly, then reconnect 

Device defective 

Return to manufacturer for repair 

Err.1 

Measured value above allowable range 

Check: pressure not within sensor range? 
-> measuring value to high! 

Sensor defective 

Return to manufacturer for repair 

Err.2 

Measured value below allowable range 

Check: pressure not within sensor range? 
-> measuring value to low! 

Sensor defective 

Return to manufacturer for repair 

Err.3 

Display range overflow 

Check: value above 19999->to high to be 
displayed 

Err.4 

Display range underflow 

Check: display below -19999 (Tara?)-> to 
low!  

Err.11 

Value could not be calculated 

Choose different unit 

Calculation overflow happened 

Choose different unit 

Err.7 

System error 

Return to manufacturer for repair 

---- 

Sensor not present/recognised 

Connect suitable sensor 

Could not calculate value 

 
 
 

Содержание HND-P105

Страница 1: ...Operating Instructions for Handheld Pressure Measuring Devices with External Pressure Sensors Model HND P105...

Страница 2: ...uration 7 7 6 Measuring Of Water Level Display Unit m 8 7 7 Calibration Services 9 7 8 Pressure Connection To The Sensors 9 7 9 Error And System Messages 10 7 10 The Serial Interface 11 8 Maintenance...

Страница 3: ...before shipping and sent out in perfect condition Should damage to a device be visible we recommend a thorough inspection of the delivery packaging In case of damage please inform your parcel service...

Страница 4: ...upply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load voltage We therefore recommend using regulated voltage...

Страница 5: ...make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be...

Страница 6: ...uring from then on will be displayed relatively to the set tare values When the tare function is activated the arrow Tara appears in the display To deactivate tare function press Tara for 2 seconds Pl...

Страница 7: ...on 7 5 1 Unit Choice Of The Display Unit Choose the desired display unit the referring unit is displayed by means of a functional arrow in the display The selection is permanently stored in the sensor...

Страница 8: ...r deviations Input is in the display unit 7 5 6 SCAL Adjusting Sensor Scale The scale of the measuring can be influenced by this setting factor is in displayed value measured value 1 Scal 100 Standard...

Страница 9: ...l dia of 4 mm to pressure port B Port A will not be used Pressure sensors HND PS01 PS02 and P03 allow for measurements of under pressure up to the entire over pressure measuring range by re plugging t...

Страница 10: ...thout battery wrong voltage or polarity Check power supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufactu...

Страница 11: ...nge unit 7 Read max value 204 Read display range decimal point 12 Read ID number 208 Read of channels 176 Read min measuring range 214 Read Scale adjustment 177 Read max measuring range 216 Read Offse...

Страница 12: ...ower off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7mm high for measuring values and for min max memories hold functio...

Страница 13: ...0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 350 0 mbar rel 0 1 mbar max 1 4 bar 0 70 C G male stainless steel HND PS09 0 1000 mbar abs 1 mbar max 4 bar abs HND PS10 0 3500 mbar abs max 14 bar abs...

Страница 14: ...face converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with lo...

Страница 15: ...ectrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic product...

Отзывы: