Operating Instructions
for
Handheld Pressure Measuring Devices
with External Pressure Sensors
Model: HND-P105
Страница 1: ...Operating Instructions for Handheld Pressure Measuring Devices with External Pressure Sensors Model HND P105...
Страница 2: ...uration 7 7 6 Measuring Of Water Level Display Unit m 8 7 7 Calibration Services 9 7 8 Pressure Connection To The Sensors 9 7 9 Error And System Messages 10 7 10 The Serial Interface 11 8 Maintenance...
Страница 3: ...before shipping and sent out in perfect condition Should damage to a device be visible we recommend a thorough inspection of the delivery packaging In case of damage please inform your parcel service...
Страница 4: ...upply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load voltage We therefore recommend using regulated voltage...
Страница 5: ...make sure the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be...
Страница 6: ...uring from then on will be displayed relatively to the set tare values When the tare function is activated the arrow Tara appears in the display To deactivate tare function press Tara for 2 seconds Pl...
Страница 7: ...on 7 5 1 Unit Choice Of The Display Unit Choose the desired display unit the referring unit is displayed by means of a functional arrow in the display The selection is permanently stored in the sensor...
Страница 8: ...r deviations Input is in the display unit 7 5 6 SCAL Adjusting Sensor Scale The scale of the measuring can be influenced by this setting factor is in displayed value measured value 1 Scal 100 Standard...
Страница 9: ...l dia of 4 mm to pressure port B Port A will not be used Pressure sensors HND PS01 PS02 and P03 allow for measurements of under pressure up to the entire over pressure measuring range by re plugging t...
Страница 10: ...thout battery wrong voltage or polarity Check power supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufactu...
Страница 11: ...nge unit 7 Read max value 204 Read display range decimal point 12 Read ID number 208 Read of channels 176 Read min measuring range 214 Read Scale adjustment 177 Read max measuring range 216 Read Offse...
Страница 12: ...ower off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7mm high for measuring values and for min max memories hold functio...
Страница 13: ...0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 350 0 mbar rel 0 1 mbar max 1 4 bar 0 70 C G male stainless steel HND PS09 0 1000 mbar abs 1 mbar max 4 bar abs HND PS10 0 3500 mbar abs max 14 bar abs...
Страница 14: ...face converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with lo...
Страница 15: ...ectrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical and electronic product...