19
Sensors with Solid Electrolyte
SensoGate WA 132
Installing the Sensor
1) Before installing the sensor, make sure that the
retractable fitting is in
SERVICE position
(red service cap
(L)
is visible).
2) Use appropriate sensors
(A)
only:
Diameter: 12 mm
Length: 225 mm
Observe the pressure resistance of the sensor.
3) Make sure that the slide washer
(C)
and the O-ring
(D)
are correctly positioned and not damaged.
ZU 0647
C
D
V
G
A
L
J
Long Immersion Depth
4) Insert the sensor and screw in the sensor head
(J)
(19 mm A/F, PG 13.5 thread) with a max. torque of
3 Nm (recommended tool: 19 mm mounting wrench,
e.g., Knick ZU 0647).
5) Thread the cable jack
(G)
through the extension
(V)
.
6) Connect the cable jack
(G)
with the sensor plug
(connection with coupling nut).