Drive Unit
SensoGate WA 132
NOTICE!
Be sure to follow the steps below in the correct order.
Take appropriate safety precautions against escaping process fluids.
Disconnect all process media and process pressure!
1) Move the retractable fitting to the SERVICE position.
2) Make sure that no process fluid is leaking from the outlet
(I)
.
3) If required, remove the sensor as described on page 15.
4) Separate the outlet and rinse connection if required.
5) Carefully turn the coupling nut
(AH)
counterclockwise (using the ZU 0680 accessory wrench
(MS)
if required – see figure). Take care to not cant the unit!
6) Pull off the drive unit
(AD)
upwards.
I
AH
AD
MS
Removing the Drive Unit