Knick SE 555X NMSN Series Скачать руководство пользователя страница 4

Hazardous Areas: Electrical and Thermal Parameters

For Sensors with VP and S8 Connector

Certificate Number: 

Marking: 

PTB 14 ATEX 2004

 II 1/2 G   Ex ia IIC T3/T4/T6 Ga/Gb

Electrical Parameters:

 

•   From a safety perspective, all circuits including the cable shield shall be considered as galvanically 

 connected to ground.

•  Metallic process connection parts shall be included in the local equipotential bonding system.

•  The effective internal inductance L

i

 and capacitance C

are negligibly small.

For Sensors with Temperature Detector (VP Connector):

Temperature class

Maximum input  

voltage U

i

Maximum input  

current I

i

Maximum input  

power P

i

T6

12 V

30 mA

50 mW

T4

15 V

80 mA

110 mW

T3

18 V

170 mA

200 mW

For Sensors without Temperature Detector (S8 Connector):

Temperature class

Maximum input  

voltage U

i

Maximum input  

current I

i

Maximum input  

power P

i

T6, T4, T3

18 V

170 mA

200 mW

Thermal Parameters:

 

The integrity of the equipment is not impaired when the standardized atmospheric conditions are ex-

ceeded within the limits of the manufacturer's specifications, e.g., with regard to ambient temperature 

and pressure.
For Sensors with Temperature Detector (VP Connector):
The maximum heating to be considered with regard to the temperature class is consistently based on 

the available power of the externally connected circuits and is 427 K/W in air. The maximum permissible 

ambient temperature is 40 °C.

Temperature class

Maximum input power P

i

Maximum permissible process temperature

T6

50 mW

55° C

T4

110 mW

80 °C

T3

200 mW

110 mW

100 °C

135 °C

 

For Sensors without Temperature Detector (S8 Connector):
The maximum heating to be considered with regard to the temperature class is negligibly low.  

The maximum permissible ambient temperature is 60 °C.

Temperature class

Maximum input power P

i

Maximum permissible process temperature

T6

200 mW

80° C

T4

200 mW

130 °C

T3

200 mW

135 °C

Special Conditions

None

For Sensors with Memosens Connector

Certificate Number: 

Marking: 

BVS 16 ATEX E 037 X

IECEx BVS 16.0030X

  II 1G 

Ex ia IIC T3/T4/T6 Ga

Thermal Parameters:

Temperature class

Ambient temperature range Ta Maximum permissible process temperature

T6

-20 °C < Ta < +70 °C 

70 °C

T4 

-20 °C < Ta < +120 °C 

120 °C

T3

-20 °C < Ta < +135 °C

135 °C

Special Conditions

•   The cable and the sensor shall only be used within the ambient temperature range specified for the 

temperature class.

•   The measuring cable including its connecting head must be protected from electrostatic charging 

if it passes through areas of Zone 0 (category 1G).

•   The Memosens sensors shall not be operated in electrostatically critical processing conditions.  

Intense vapor or dust flows directly impacting on the connection system shall be avoided.

•   Metallic process connection parts must be mounted at the installation site so that they are  

electrostatically conductive (< 1 MΩ). 

Содержание SE 555X NMSN Series

Страница 1: ...urch elektrostatische Aufladung stark erhöhen Entfernen Sie gegebenenfalls Luftblasen im Raum hinterdempH sensitivenGlasdurchleichteSchleu derbewegungen in senkrechter Ebene Setzen Sie den Sensor in die Armatur ein wie in der Anleitung der Armatur beschrieben Verbinden Sie Sensor und Kabel miteinander 4 Betrieb 4 1 Kalibrierung des Sensors Für den Sensor SE 555 wird eine 2 Punkt Kalibrie rung empf...

Страница 2: ...tromkreise und beträgt 427 K W in Luft Die maximal zulässige Umgebungstemperatur beträgt 40 C Temperaturklasse Maximale Eingangsleistung Pi Maximal zulässige Prozesstemperatur T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C Für Sensoren ohne Temperaturfühler S8 Kopf Die im Hinblick auf die Temperaturklasse maximal zu berücksichtigende Erwärmung ist vernachläs sigbar klein Die maximal zul...

Страница 3: ...since this can lead to electrostatic charging and sluggish response times Check the space behind the pH sensitive glass for the presence of any air bubbles and remove them by gently shaking the sensor up and down Please refer to the user manual of the respective fitting for instructions on how to install the sensor Connect sensor and cable 4 Operation 4 1 Calibrating the Sensor 2 point pH calibrat...

Страница 4: ...y connected circuits and is 427 K W in air The maximum permissible ambient temperature is 40 C Temperature class Maximum input power Pi Maximum permissible process temperature T6 50 mW 55 C T4 110 mW 80 C T3 200 mW 110 mW 100 C 135 C For Sensors without Temperature Detector S8 Connector The maximum heating to be considered with regard to the temperature class is negligibly low The maximum permissi...

Отзывы: