24
Measurement
cal
CAL
CALIMATIC blinks
Use
to select the required calibration
method (CALIMATIC, MANUAL, DATA INPUT,
ISFET‑ZERO, CAL SOP (Option 001), ORP OFFSET
(for ORP or pH/ORP combo sensors), or TEMP.
OFFSET (Option 001)).
cal
Perform the selected calibration
as described on the previous pages.
The meter then automatically returns to measuring mode.
FREE CAL Calibration
(Free selection of calibration method)
FREE CAL calibration is selected in the configuration menu.
Note:
To abort calibration, you can press
meas
at any time.
Calibration
Содержание PORTAVO 904X PH
Страница 1: ...www knick de Read before installation Keep for future use Portavo 904 X PH Portable Meter User Manual ...
Страница 2: ......
Страница 50: ...50 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Страница 51: ...51 pH 02 Knick CaliMat Nominal values in bold C pH Buffer Tables ...
Страница 52: ...52 pH 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated Buffer Tables ...
Страница 53: ...53 pH 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added Buffer Tables ...
Страница 55: ...55 pH 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH Buffer Tables ...
Страница 56: ...56 pH 07 WTW Technical Buffers Nominal values in bold C pH Buffer Tables ...
Страница 57: ...57 pH 08 Hamilton Nominal values in bold C pH Buffer Tables ...
Страница 58: ...58 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Страница 59: ...59 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...
Страница 66: ...66 V Value added features 8 Viewing recorded data 32 Viewing the logger data 32 Index ...
Страница 67: ......