22
The Data Logger
The meter provides a data logger. Prior to use, it must be configured and then
activated. You can choose from the following logger types:
• DIFF (signal-controlled logging of measured variable and temperature)
• INT (time-controlled logging at a fixed interval)
• DIFF+INT (combined time- and signal-controlled logging)
• SHOT (manual logging by pressing the STO key)
The data logger records up to 5000 entries and saves them in a circular buffer.
Already existing entries will be overwritten.
The following data are recorded: primary value, temperature, time stamp and device
status.
The Paraly SW 112 software allows convenient management of the data logger.
It is always the currently selected process variable which is recorded.
The “STO” icon and the memory address is displayed briefly to indicate that an entry
is being saved.
Data Logger
RCL:
Measured
values are
displayed
Logger type
Date
Data logger activated
Primary
process
variable
Temperature
Display: Icons related to the data logger
Measured
value is being
saved: STO
Memory
address
(0026)
Next value
Time