pH Configuration
Prior to measurement, a configuration should be performed to match
the connected sensor and the desired measurement performance.
Furthermore, you can select the suitable calibration method.
The following table gives you an overview. Factory settings are
shown in bold print.
Measurement
“Setup” display
set
Select using arrow keys, confirm by pressing set.
Display 1
pH x.xx / pH x.xxx / mV /
(°C for analog pH only)
Display 2
OFF / date + time / date / time
CAL Timer
OFF / 1 ... 99 days
CAL
CALIMATIC/Manual/DATA INPUT/
(ISFET-Zero)
/FREE CAL
CAL POINTS
1 / 2 / 3 / 1-2-3 (for CALIMATIC, Manual, FREE CAL)
BUFFER SET
(CALIMATIC,
FREE CAL)
-01- Mettler Toledo
2.00/4.01/7.00/9.21
-02- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-03- Ciba (94)
2.06/4.00/7.00/10.00
-04- NIST technical
1.68/4.00/7.00/10.01/12.46
-05- NIST standard
1.679/4.006/6.865/9.180
-06- HACH
4.01/7.00/10.01/12.00
-07- WTW techn. buffers 2.00/4.01/7.00/10.00
-08- Hamilton
2.00/4.01/7.00/10.01/12.00
-09- Reagecon
2.00/4.00/7.00/9.00/12.00
-10- DIN 19267
1.09/4.65/6.79/9.23/12.75
Auto OFF
OFF/ 0.1h / 1h / 6h / 12h
Temp Unit
°C / °F
Time Format
24h / 12h
Date Format
dd.mm.yy / mm.dd.yy
Default
NO / YES (reset to factory settings)
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing set.