18
Specifications
Connections
1 x pH socket, DIN 19 262
2 x 4-mm socket for separate temperature detector
1 x M8 socket, 4 pins, for Memosens lab cable
Display
LCD STN 7-segment display with 3 lines and icons
Sensoface
Status indication (friendly, neutral, sad)
Status indicators
Battery power level
Notices
Hourglass
Keypad
[on/off], [cal], [meas], [set], [
▲
], [
▼
], [clock]
Diagnostics functions
Sensor data
(Memosens only)
Manufacturer, sensor type, serial number, operating time
Calibration data
Calibration date, zero, slope
Device self-test
Automatic memory test (FLASH, EEPROM, RAM)
Device data
Device type, software version, hardware version
Data retention
Parameters, calibration data > 10 years
EMC
EN 61326-1 (General Requirements)
Emitted interference
Class B (residential area)
Immunity to interference
Industry
EN 61326-2-3
(Particular Requirements for Transmitters)
RoHS conformity
According to directive 2011/65/EC
Power supply
Portavo 902
4 x AA alkaline batteries
Operating time
Approx. 1000 h (alkaline)
Nominal operating
conditions
Ambient temperature
-10 … +55 °C
Transport/
Storage temperature
-25 … +70 °C
Relative humidity
0 … 95 %, short-term condensing allowed
Housing
Material
PA12 GF30 (silver gray RAL 7001) + TPE (black)
Protection
IP 66/67 with pressure compensation
Dimensions
Approx. (132 x 156 x 30) mm
Weight
Approx. 500 g