Calibrating
Measurement
The number of calibration points has been
selected in the configuration menu.
pH display blinks
Use
▲▼
to set the temperature-corrected pH
value taken from the buffer table.
mV display blinks
Depending on the number of calibration points,
the procedure described above for CAL 1/2/3 is
repeated.
mV value blinks until calibration is completed, then successive display of:
Then the meter switches to measuring mode.
MANUAL Calibration
(Manual calibration)
The calibration method is selected in the configuration menu.
Note: To abort calibration, you can press meas at any time. This will be confirmed
by the display message “CAL ABORTED”. Exception: When you have selected “CAL
POINTS 1-2-3” and the first calibration step has been completed, the calibration
process cannot be stopped any more.